EG&G ORTEC 463 Manual de funcionamiento y servicio - Página 18
Navegue en línea o descargue pdf Manual de funcionamiento y servicio para Instrumentos de medida EG&G ORTEC 463. EG&G ORTEC 463 20 páginas. Constant fraction discriminator
12
produce two or more discriminator output pulses. In the
successfully on even longer decay scintillators, but the
463 this problem is overcome by using the Nal mode in
internal dead time may have to be increased to prevent
which an internal dead time of 'vl jisec is generated. The
multiple triggering; to increase the dead time, increase
463 can then be operated in the Nal mode and be used
the capacitance of C4.
BIBLIOGRAPHY
THEORY OF CONSTANT FRACTION OF PULSE HEIGHT
DISCRIMINATORS
1. D.A. Gedcke and W.J. McDonald, "A Constant Fraction
of Pulse Height Trigger for Optimum Time Resolution,"
Nucl. Instr. Methods 55, 377 (1967).
2. D.A. Gedcke and W.J. McDonald, "Design of the
Constant Fraction of Pulse Height Trigger for Optimum
Time Resolution," Nud. Instr. Methods 58, 253 (1968).
3. W.J. McDonald and D.A. Gedcke, "Electronics for Fast
Neutron Work," p. 56 in International Symposium on
Nuclear Electronics, Versailles, September 1968, Vol. 1.
TIMING WITH Ge(Li) DETECTORS
4. R.L. Chase, "Pulse Timing System for Use with Gamma
Rays on Ge(Li) Detectors," Rev. Sci. Instr. 39{9), 1318
(1968).
5. R.L. Graham, I.K. MacKenzie, and G.T. Ewan, "Timing
Characteristics of Large Coaxial Ge(Li) Detectors for
Coincidence Experiments," IEEE Trans. Nud. Sci. NS-13
(1), 72 (1966).
TIMING WITH SILICON DETECTORS
6. A. Alberigi Quaranta, "On the Information Available
from the Rise-Time of the Charge Pulse Supplied by a
Semiconductor Particle Detector," Nud. instr. Methods
35, 93 (1965).
7. M. Moszynski and B. Bengtson, "Plasma Delay and
Plasma Time Jitter in Subnanosecond Timing with Surface
Barrier Detectors," Nud. Instr. Methods 91, 73 (1971).