Ega Master 51263 Manual de instrucciones - Página 7
Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida Ega Master 51263. Ega Master 51263 16 páginas. Sound meter
CALIBRATION (see figure 4)
1. Turn the Meter on.
2. Put the Meter in the "A" weighting mode, FAST
response mode, range set to 60~110dB, lock to MAX.
3. Place the microphone onto the calibrator's 1/2 inches
sound source hole.
4. Turn the calibrator on, using 94dB@1kHz standard
sound source.
5. Adjust the Meter's CAL potentiometer located on the
side panel until the LCD displays 94.0dB
figure 4
SPECIFICATIONS
A. General Specifications
Display: 3 1/2 digits, 1999 maximum
Overloading:
Under range displays UNDER
Over range displays OVER
Battery Deficiency:
Change batteries as soon as
is displayed.
Sampling Rate:
Fast Speed: 125 microseconds
Slow Speed: 1 second
Microphone: 1/2" electret condenser
7