Ega Master 57665 Manual de instrucciones - Página 10

Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida Ega Master 57665. Ega Master 57665 18 páginas. Máquina de calibración de par

Ega Master 57665 Manual de instrucciones
1
.
Turn the meter on and select the desired torque.
2
.
Place the torque wrench in the housing.
3
.
We turn the crank to the left until the torque shown by the meter drops to the
desired torque (if you want to raise the torque, turn the crank to the right).
1
Remark: For making correct calibration test, it is advised to follow international
standard procedures: for example, international standard ISO6789 or similar.
Important: Overload in the toque tester can damage it.
Keep the machine clean.
Perform periodic torque tester calibration.
10
QUICK USE GUIDE
Before starting the calibration:
-
the torque wrench must not be loaded.
-
make sure the torque tester is on when calibrating.
2
MAINTENANCE
3