7.
Connect the rapid inflation pump to the CPR.
Keep the CPR and pump in the same axis in order to avoid arching the CPR by the
weight of the pump. Incorrect use could cause a water tightness defect for the
CPR.
8.
Press the ON button for the rapid inflation pump until the support is inflated
(approx. 1 minute) - Note: The level of inflation has no significance because it will
be adjusted automatically by the compressor.
9.
Close the CPR valve by positioning the index finger on CLOSED,
10. Connect the support to the compressor.
11. After the hourglass LED has gone off, the patient can be installed on the support.
NOTES:
 The information needed to use the compressor is on the side of it: simplified leaflet.
 After 5 minutes without a selection, the brightness of the LEDs is lessened to avoid
disturbing the user at night. Pressing the key reactivates maximum brightness for 5
minutes.
 The pneumatic connector has a stopper to re-equilibrate pressure of the support
when it is disconnected from the compressor.
WARNING:
 The electrical cable must be installed in such a way as to avoid any severing with
the articulated parts or the wheels of the bed and prevent staff from stumbling. Risk
of physical injury and equipment damage. Use the removable attachment systems
provided.
 The power supply cable is the severing system for the machine.
 The compressor keyboard, the power supply cable, the pneumatic connector and
the CPR valve must be permanently visible and accessible.