CO2 Meter RAD-0002 Manuel d'utilisation - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} CO2 Meter RAD-0002. CO2 Meter RAD-0002 14 pages. Oxygen monitor, remote oxygen storage safety system
Également pour CO2 Meter RAD-0002 : Manuel d'utilisation (2 pages)

CO2 Meter RAD-0002 Manuel d'utilisation

1.8 Recover Factory Settings

If calibration or changing custom parameters has caused the RAD-0002 to behave incorrectly,
use the ReFactSet function to return the unit to its original condition at shipping.
1. Press Mode until the "ReFactSet" icon flashes.
2. Press Up/Down button to choose either "Yes" or "No".
3. Press Enter to save or press Mode to exit without saving.

11. Product Care

To insure you receive the maximum benefit from using this product, observe these guidelines:
Repair - Do not attempt to repair the product or modify the circuitry by yourself. Please
contact CO2Meter directly if the product needs servicing, including the replacement or
calibration of sensor See section 15 for technical support contact information.
Cleaning - Disconnect the power before cleaning. Use a damp cloth. Do not use liquid
cleaning agents such as benzene, thinner or aerosols, as these will damage the device.
Maintenance – We recommend testing the communication between the SEU and RDU
under 'DIAG" function. If these five LEDs blink and the buzzer of SEU and RDU sound
simultaneously, it indicates that SEU and RDU work normally.

12. Safety Notes

Warning: Your safety is very important to us. To ensure to use the product correctly and safety,
please read these warnings and the entire User Manual before using the product. Otherwise,
the protection provided by the equipment may be impaired. These warnings provide important
safety information and should be observed at all times.
1. Please handle the device carefully; do not subject the product to impact or shock.
Otherwise, this may cause the accuracy drift.
2. Do not place the unit near a heat source. Heat can cause distortion of the unit, which may
result in an explosion or fire.
3. Do not touch the exposed electronic circuitry of the device under any circumstances, as
there is the dangerous of electric shocks.
4. Please take care of cable connection between SEU and RDU. Make sure the cable from
SEU is connected into the INPUT port of RDU.
5. Please ensure the external power supply is normally supplied to ventilation fan while the
relay is working. If there has no normally power supply to the fan, the relay will not work,
which may result in potential danger with lower O2 concentration in confined space.
10