5.2 SETTING THE CH1 OR CH2 OUTPUT CURRENT

1. Turn to the power supply
2. Refer to Section 5.1 step 1-4 to give the voltage around 2-5V
3. For CH1, turn the CH1 CURRENT TUNE KNOB [19] anticlockwise reach the minimum current value.
4. Short the + and the - MAIN TERMINAL [24, 25] with a conductor which is cross section area not less than
2
0.5mm
.
5. Ensure the output indicator is on. Otherwise, press the AUTO CURRENT CUT OFF PROTECTION KEY
[8].Then the CH1 CV/CC INDICATOR [17] will turn to red color.
6. Adjust the CH1 CURRENT KNOB [19] to give a desired output current.
7. Repress the AUTO CURRENT CUT OFF PROTECTION KEY [8] to Cut off the output.
8. Then the CH1 CV/CC INDICATOR [17] will turn to green color.
9. Remove the conductor the MAIN TERMINA [24,25]
10. Set to desired voltage.
11. Connect the circuit to the MAIN TERMINA [24,25]
12. To setting CH2 current, repeat the above steps, use CH2 CURRENT TUNE KNOB [4], short MAIN TERMINAL
[26, 27] and CH2 CV/CC INDICATOR [6] instead.
Remarks:
The conductor is not provided
Caution:
l
Ensure the current is set to zero before shorting the MAIN TERMINALS. Otherwise it will damage the power
supply.
l
Do not short the MAIN TERMINALS over 1 minute; it will damage the power supply.