Astrea Evolve AA25CRK Manuel de remplacement - Page 9
Parcourez en ligne ou téléchargez le pdf Manuel de remplacement pour {nom_de_la_catégorie} Astrea Evolve AA25CRK. Astrea Evolve AA25CRK 19 pages. Wetted flow path for 200mm and 250mm chromatography columns
10. Discard the adjuster bed support, flowcell and antijet.
Wetted Flowpath components Discarded
Flowtube adaptor
Flowtube seals x2
Flowtube
Adjuster mesh assembly
Adjuster Flowcell and antijet
Adjuster seal
Disassembly of the Fixed Assembly
1. Identify the components on the schematic shown in Figure 1.3.
2. Disconnect the pipework elbow from the bottom of the column.
3. Identify the fixed end support blocks that are located on the column support legs. Swing
these blocks so that they support the flowcell. These will support the flowcell when the
fixing bolts are removed. See Figure 1.3.
ADJUSTER ASSEMBLY
Figure 1.3 Fixed Assembly Evolve® 250 column
Non-wetted components RETAINED
Seal Actuation Nut
Spacer
Stainless steel Seal adjust tube
Threaded central height adjust rod
Height adjust handle assembly
Adjust Flange
Seal setting ring
Flowcell locking clip
8