ASC-400 Adjustment procedures

5.1

Adjusting Isolated current (mA) Input:

1. Connect Calibrator mA output to Digital Multimeter mA input and to ASC-400
Isolated mA measure terminals.
2. Type "1" in the Calibration menu.
Prompt = "Input 0 mA to ASC400 - Press <enter> when stable"
3. Apply 0.000 mA (Digital Multimeter reading) to the input and press "enter"
when stable.
Prompt = "Input 24 mA to ASC400 - Press <enter> when stable"
4. Apply 24.000 mA (Digital Mulimeter reading) to the input and press "enter"
when stable. ASC-400 now returns to the Calibration menu.
5.2

Adjusting Isolated voltage Input:

1. Connect Calibrator V output to Digital Multimeter V input and to ASC-400
Isolated V measure terminals.
2. Type "2" in the Calibration menu.
Prompt = "Input 0 Volt to ASC400 - Press <enter> when stable"
3. Apply 0.000 V (Digital Multimeter reading) to the input and press "enter" when
stable.
Prompt = "Input 30 Volt to ASC400 - Press <enter> when stable"
4. Apply 30.000 V (Digital Multimeter reading) to the input and press "enter"
when stable. ASC-400 now returns to the Calibration menu.
5.3

Adjusting current (mA) Input:

1. Connect Calibrator mA output to Digital Multimeter mA input and ASC-400 mA
measure terminals.
2. Type "3" in the Calibration menu.
Prompt = "Input 0 mA to ASC400 - Press <enter> when stable"
3. Apply 0.000 mA (Digital Multimeter reading) to the input and press "enter"
when stable.
Prompt = "Input 24 mA to ASC400 - Press <enter> when stable"
4. Apply 24.000 mA (Digital Multimeter reading) to the input and press "enter"
when stable. ASC-400 now returns to the Calibration menu.