EGA-501CO2 Insuf ator
Insuf ator Front Panel:
1. On/Off Button functions properly and actuates to turn pump on/off.
2. H20 button remains green to indicate all heating capabilities are functioning properly. It will turn ash blue if the heater
is unplugged.
3. CO2 button remains green to indicate all CO2 functions are connected properly. It will turn blue if the ow is off or the
pressure source is not connected.
4. Flow rate knob should move freely without hesitation.
5. H, M, L letters should light up with each setting.
6. Push the knob OR turn past the L setting to stop the ow of CO2.
7. "Low Pressure" is ashing when CO2 source is turned off or not attached.
Bottle Warmer:
1. Securely fastened to the base of pump
2. Connection is secure and heats properly.
3. Bottle ts securely in holder once wing nut is tightened.
SYSTEM SETUP
EGA-500 Irrigation Pump
1. Attach the bottle holder and base to the device; plug the heater cable into the rear of the device.
2. Plug the power cord into the rear of the device and set the device on the endoscopy cart, ensuring that there is clearance to
insert and remove bottle.
3. Attach the pump head and foot pedal to the front of the device.
EGA-501 CO2 Insuf ator
1. Attach the bottle holder and base to the device; plug the heater cable into the rear of the device.
2. Plug the power cord into the rear of the device and set the device on the endoscopy cart, ensuring that there is clearance to
insert and remove the bottle.
3. Attach input hose to the rear of the device and then to the source (tank or wall).
OPERATOR CONTROL
EGA-500 Irrigation Pump
1. Flow rate control
2. Foot pedal activation
3. Prime feature. It can be stopped by opening the pump head, pressing the button, or pressing the foot pedal.
4. Loading pump head. Motor will not turn while the pump head is open.
5. Water warming. Prewarming is required.
EGA-501 CO2 Insuf ator
1. Flow rate control. System always powers up in "off". Pressing knob stops ow. Flow rates imitate Olympus light source's
air pump.
2. Low pressure light: When ashing, ow will still continue until the tank is fully depletedis not stopped.
3. Water warming: Prewarming is required.
4. CO2 warming: The heater is disabled (blue steady light) when the ow is turned off or the low pressure light is on.
SYSTEM OPERATIONS
EGA-500 Irrigation Pump
Startup
1. Verify power is on.
2. Turn on water warmer if desired.
Daily Service
1. Load tube set (EndoGator® or Hybrid)
2. Press the prime button to prime the endoscope
Pre-procedure
1. Prime the endoscope by pressing the prime button.
Procedure
1. Adjust ow rate if needed.
2. If the bottle needs to be replaced, follow instructions on tube set IFU regarding use of sterile technique.
End of Day Shutdown
1. Dispose of bottle and tube set.
2. Turn off power.
EGA-501 CO2 Insuf ator
Startup
1. Verify power is on.
2. Turn on water warmer if desired.
3. Turn on CO2 source; verify "Low Pressure" light turns off.
Daily Service
1. Connect 100551 Endo SmartCap® CO2 source tubing to CO2 outlet and Endo SmartCap® or EndoGator® Hybrid
according to 100551 and Endo SmartCap® (or Hybrid) directions.
Preprocedure
1. Turn on CO2 ow according to desired ow and turn on CO2 heater if desired.
2. Prime endoscope lens rinsing water channel and verify insuf ation gas by bubbling.
3. Note that warming a bottle from room temperature to body temperature will take over an hour. Water bottles should be
prewarmed.
4. CO2 won't ow when display indicates L, M, or H status
5. Check for adequate source pressure. Be sure that the source's valve is open.
6. Check to see whether the low pressure light is on. This would indicate that there is insuf cient source pressure.
7. Check for clogs in the CO2 output tubing ow path.
CO2 isn't warm
1. Check the temperature at the output of the device; it is easier to detect the difference at that point.
2. Note that the CO2 is only warmed to body temperature. It may not feel noticeably "warm" to some people, depending on
sensitivity.
3. A solid blue activation switch when the feature is rst turned on indicates that the Low Pressure light is on or the ow rate
is turned off. Ensure suf cient source pressure is present. Ensure that the ow is turned on.
End of Day Shutdown
1. Dispose of bottle and tube set.
2. Turn off power.
3. Turn of CO2 gas source.