Dagan IX2-700 Manuel d'utilisation - Page 12

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Dagan IX2-700. Dagan IX2-700 14 pages. Dual intracellular preamplifier

IX2-700 Dual Intracellular Preamplifier

SPECIFICATIONS

Electrode Holders:
Headstage:
Current
Multiplier
(N)
Max I
10
10µA
1
1µA
.1(Standard)
100nA
.01
10nA
.001
1nA
Extracellular Recording:
Filter Response:
Input Bias Current:
Rise Time to Pulse:
Voltage Gain:
Output Impedance:
Offset Voltage:
Low noise polycarbonate in two sizes: 1.0 to 1.2mm
(#HBI2O); 1.5 to 1.8 mm (#HB18O) (order separately)
Box Probe, 1.5" x 1.2" x .88", 4" long, 3/8" rod, BNC
Input Connector
Standard Headstage: N=.l for 20 to 500 Megohm
electrode
Feedback
Resistor
Noise
R
RMS/BW/Re
E
1MΩ
25µv, 10kHz, 1MΩ
10MΩ
50µv, 10kHz, 10MΩ
100MΩ
50µv, 1kHz, 100MΩ
1GΩ
40µv, 100Hz, 1GΩ
10GΩ
35µv, 10Hz, 10GΩ
Works with 8024 voltage, gain 1000 AC
coupled headstages
Switch selectable DC to 1kHz, 3kHz, 5kHz,
10kHz, 30kHz and Wide Band
1 x 10
Typically 10 microseconds into 10 megohm
resistor (10 to 90%), 50 microseconds into
100 megohm
X10 and X100
500 ohms all outputs
+200mV
Electrode
Resistance
Range
.2 to 5 MΩ
2 to 50 MΩ
20 to 500 MΩ
200 to 5000 MΩ
2 to 50 GΩ
12
-
Amps (Adjustable to zero)
11
Input
Resistance
Cp
10
20 pf
10
MΩ
11
10 pf/35 pf
10
MΩ
12
10 pf/35 pf
10
MΩ
13
20 pf
10
MΩ
14
20 pf
10
mΩ