There are bubbles on the inner wall of the measuring tank, which
will cause the measurement value to be wrong or rise continuously.
It is suggested to increase the water inflow and reduce the water
outflow so as to generate pressure in the measuring pipe and make
the bubble burst。
Due to different manufacturers and different components of
sewage, colored substances adhere to the inner wall of the
measuring pipe, and the cleaning interval and frequency depend
on the specific situation of the site。
Keep the ventilation and dry environment, condensation will
occur in the humid environment, and the adhesion to the outer
wall of the measuring tank will cause the measurement value to
increase。
Do not disassemble the right measuring tank without the technical
support of the manufacturer。