Carbolite 200 Manuel d'utilisation - Page 4
Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Carbolite 200. Carbolite 200 8 pages.
200 Controller
3
O
VERTEMPERATURE
This is only fitted when the overtemperature option is ordered. To identify whether it is fitted,
observe the controller panel: if it is fitted the overtemperature key and warning light are present.
The overtemperature key must be held depressed continuously while viewing or adjusting the
overtemperature controller. If the Overtemperature key is released, the display reverts to the main
controller.
With the key pressed the overtemperature setpoint is normally displayed. To alter the value, use
Down and Up. To view the measured temperature, press Scroll until
displayed, then press Down or Up. Keep the overtemperature key depressed during these
operations.
When an overtemperature condition is reached, the overtemperature warning light changes from
green to flashing red. To "accept" or "reset" this alarm condition, press the Overtemperature key
and the Page key together; the warning light changes to continuous red. Once the over-
temperature condition is cleared (by the temperature falling), the light changes to green and
normal operation resumes.
Audible Alarm (option only)
If an audible alarm is supplied for use with the overtemperature condition, then it is normally
configured to sound on overtemperature, and to go off when the alarm is acknowledged as
described above.
It is not possible to cover in this manual other possible alarm features which may be included by
customer special order.
4
P
L
OWER
IMIT
There is no power limit parameter in the model 200.
Any reference to the power limit in the oven manual can be disregarded. If a power limit is
required, the 200 cannot be used.
4
O
PTION
PV
(process variable) is
MC00 – 1.03