FOLLO FUTURA Stabilis Series Manuel - Page 8

Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} FOLLO FUTURA Stabilis Series. FOLLO FUTURA Stabilis Series 18 pages.

FOLLO FUTURA Stabilis Series Manuel
F
S
W
OLLO
TABILIS
ALKER
Unpacking/installation
1.
Make sure that the product has not been damaged (hurt) in
transport.
2.
Fit the walker according to assembly instructions.
3.
Assemble all the accessories that have been ordered.
4.
Test all the various functions and setting options.
Security information
1.
If there is any damage or defects on the walker, it should not
be used until these are rectified
2.
The Walker is designed for indoor use only.
3.
When adjusting height and depression, make sure that
nothing is pinched between the frame parts.
4.
The Walker should never be charged with more weight than
the approved static user weight.
5.
Infusion stand has to be mounted parallel to the walker (see
illustration on page 16).
6.
The electric walker should not be used with the charger
attached to the wall outlet.
Warrenty
The product must not be modified or used with other items than
original parts. The manufacturer responsibility for CE marking is
invalidated if any form of modification is carried out.
Follo Futura provides 2 year warranty on materials and
manufacturing errors, provided that the product is used according
to the product's purpose and according to the instructions.
Surface treated parts, pads, wheels, brake parts or handles are
not included.
This product has been assembled, controlled and approved in
accordance with the company's quality assurance system.
Adjustment
Height control of walker with gas spring: The handle at the front
of the top frame activates the gas spring. Squeeze the handle
towards the top frame to adjust the walker up or down. To lower
the walker, squeeze the handle and load the armrest with body
weight, then drop the handle at the desired level.
Height control of walker with electric motor: The height of the
armrests is adjusted with the controller to the desired height. The
product is also delivered with manoeuvring feature integrated into
the handles.
Height adjustment of the manual version:
Height is adjusted by loosening the clamp
handle on the front tube. Adjust the top
110
section to the correct height and tighten the clamp handle.
Make sure the clamp handle is securely tightened.
Cushion adjustment:
Loosen the handles. Adjust the
cushions to the desired position,
tighten the handles (make sure they
are tightened). Please note: Support
handles follow the armrest setting and
facilitates appropriate support and
use.
Locking the wheels: The rear wheels have locks. To unlock the
brakes, press down the "locking" with the foot. To release the
brakes, tilt "locking" up again with the toe.
Maintenance and inspection
Control at regular intervals (for example every third month):
1.
All screws are tight and all functions are working properly.
2.
There is no damage or wear and tear on the walker. The
walker must be stable.
3.
The braking effect on the brakes is adequate.
4.
The telescopic tubes are lubricated regularly (with a thin
layer of, for example, vaseline or silicone).
5.
The cushions are made of polyurethane foam, and can
easily be wiped off with most detergents. Attention!
Alcohol-based products may discolour the cushions and
should not be used. The motor, the control box and the
controller should be wiped off with a slightly moist cloth.
Renovation and preparation
Renovation and preparation of the product must be
performed by professionals.
Before cleaning/disinfection, remove the cushions. Control
box is attached with velcro and may be easily removed from
the base frame, the hand control follows. Engine
disassembles easily by removing two M10 bolts and nuts,
see the groundbreaking design.
After cleaning/disinfecting, wipe product carefully so that
moisture dislodges. It is very important that all cavities are
inspected for moisture.
When the product is dry, telescopic tube must be lubricated
with silicone spray and top frame is moved up and down a
few times so that the parts slide easily. After mounting of
cushions, check the sideway adjust alignment and if the felt
is in place.
Replace broken, and worn parts by ordering them from the
supplier, see the exploded drawing and part list.
All worn/damaged information- and warning labels must be
replaced. The labels must be ordered from the supplier.
The product assembles in accordance with groundbreaking
design.
Recycling
The motor, the control box and the hand control are disposed of
as electronic waste. The cushions are disposed of as combustible
waste, while the stand etc. are sorted as metal.
Charge Routines
When charging the Follo Stabilis Walker, the charger should
always first be connected to the walker before plugging into the
wall outlet. Disconnection is performed in the reversed order. The
batteries are usually fully charged at delivery, but they should
nevertheless be set to charge overnight before Follo Stabilis
Walker is used. The charging should always be done in room
temperature. It should always be charged when the warning
sound appears (a beep sound). The battery should never be
completely emptied. The battery will not be destroyed by too
much charge. NOTE: The walker should NOT be used with the
charging device (A) connected. The charging takes place via
the battery pack (B) or via the controller (C). The charging unit (A)
is inserted into a hole on the left side products. The unit is
charged via a common wall outlet.
22/12/11
Side 8
Rev: 02