Folyon H300P Manuel de l'utilisateur - Page 13
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Folyon H300P. Folyon H300P 17 pages. Temperature humidity chamber with watlow pm plus controllers
Global Menu
Communications Menu
Restoring Factory Default Parameters
Both the temperature and humidity controllers were set up at the factory for optimum operation for each
H300 chamber. These default parameters were stored in User Settings 1. To restore factory default
parameters for either controller follow the procedure below:
Press the Forward Arrow key once. The PM Plus controller will display a menu.
•
Press the – key until Setup is highlighted.
•
Press the Forward Arrow key once. The PM Plus controller will show the Setup Menu.
•
Press the Forward Arrow key until Global is highlighted.
•
•
Press the Forward Arrow once.
•
Press the – key until Restore Setting From is highlighted.
•
Press the Forward Arrow key once.
•
Press the – key until User Set 1 is highlighted.
•
Press the Forward Arrow key to restore factory defaults.
H300P Temperature Humidity Chamber
Parameter
Display Units
Language
AC Line Frequency
Comms LED Action
Zone
Channel
Display Pairs
Display Time
Bluetooth
Home Screen
Parameter
Standard Bus Address
Data Map
Non Volatile Save
Temperature
Humidity
C or F
English
English
60Hz
60Hz
Both
Both
Off
Off
On
On
1
1
0
0
On
On
0
0
Temperature
Humidity
2
1
1
1
Yes
Yes
Page 13