Quick Use of the X-123SDD
1. Connect the X-123SDD and PC as described in the previous section.
2. Launch the ADMCA software by opening the ADMCA.exe file.
3. When the "Starting ADMCA" box appears as in figure 2, select DP5 and
click on "Connect." The X-123SDD unit should now be connected to the
software. To verify this, confirm that a serial number for the DP5 is
shown in the top right corner of the software and that the USB symbol
located at the bottom right corner of the software is green.
4. Now that the X-123SDD is connected the proper configuration must be
selected for the detector. The ADMCA software includes configurations
for most Amptek detectors. These configurations are identified
by the detector material, dimensions, and cooler type.
o NOTE:
The
configuration, so when powering the system for the first
time and clicking Connect, the factory configuration is
loaded. This is the same configuration that is displayed on
test sheet shipped with the detector.
5. To select a configuration select "DPP Setup" under the "DPP"
dropdown list in the software or by clicking the "acquisition
setup" button on the toolbar. The dialog box shown in Figure 3
will appear. The configurations can be selected through the
dropdown
menu
Configuration."
6. Once the correct configuration is selected from the dropdown
menu the indented grey area above the dropdown menu should
read (for example) "Loaded SDD 7mm2/500um 2-Stage Cooler
configuration." Click "Show Current Configuration" to view
the settings. A box will appear with the settings. Click OK to
exit the box.
7. To apply the selected configuration click the "Apply" button.
Then select the "OK" button to exit this dialog box.
8. Now that the X-123SDD has loaded the appropriate
configuration for that detector, an acquisition can be started.
Place a source in front of the detector. To start an acquisition,
press the space bar. The space bar will also stop an acquisition.
It may take the detector a few minutes to stabilize, so if the acquisition does not look correct wait a few minutes and then
press the "A" key on the keyboard to clear the acquisition and begin a new one. It can take up to 2 minutes for an Amptek
detector to stabilize after a configuration has been sent.
o Once the detector has stabilized adjust some of the
thresholds which prevent low-end noise and other
unwanted counts from affecting the spectrum. This can
be done automatically by the software by clicking the
"Tune Slow/Fast" button on the toolbar. This must be
done with no source in front of the detector.
9. A common adjustment is to change the gain of the X-
123SDD. This can be done in the "Gain & Pole Zero" tab
of the DPP properties dialog box or by using the gain
buttons on the toolbar.
thresholds whenever the gain is changed. This can be done
automatically as explained in the previous step.
X-123SDD
remembers
labeled
"Read
Amptek
It is necessary to readjust the
the
last
Detector
Figure 3: Properties dialog box.
Figure 4: Acquisition spectrum for an
X-123SDD Quick Start Guide Rev A1
Figure 2: Starting ADMCA
55
Fe source.