Amptek COOL-X Manuel d'utilisation - Page 8
Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Amptek COOL-X. Amptek COOL-X 8 pages. X-ray generator with pyroelectric crystal
AMPTEK INC.
COOL-X Operating Manual
APPENDIX A
Operation of the COOL-X away from its Controller
The Amptek COOL-X, x-ray generator is designed to be used as an integrated unit together with
o
o
its Controller in normal room temperature applications 20
C ±5
C. Operating the unit outside
these temperature conditions may result in faulty operation or damage to the COOL-X emitter.
In normal operation, the COOL-X cycles through heating and cooling phases. Heat-sinking of
the emitter head (TO-8 package) to provide the proper cooling profile is accomplished by having
good thermal contact with the COOL-X Controller housing. Normal air convection is usually
adequate to dissipate the heat from the housing, but performance can be improved by attaching
the housing to a metal plate with 4 screws.
Operation of the COOL-X emitter (TO-8) away from its Controller must be done very carefully,
taking into account the following considerations:
1) Heat-sinking:
The TO-8 emitter head must be mounted on a heat-sink which is kept at room temperature. The
hole pattern for the TO-8 for attaching it to the heat-sink is provided below. The Control
electronics should also be mounted on a metal plate and kept at approximately the same
temperature as the TO-8. This is important because the temperature sensor in the Control circuit
refers to a matched temperature sensor inside the TO-8 emitter head in order to maintain the
proper temperature differentials and temperature limits throughout the heating and cooling
phases of the cycle.
2) Grounding:
The temperature sensor inside the TO-8 emitter head has a normal output voltage in the range of
0.560 to 0.410 volts corresponding to temperatures of 35 to 125°C. The temperature sensor
shares a common ground with the crystal heater which normally draws about 400 mA of current
during the heating phase. It is important to maintain a low impedance Ground return between
the TO-8 (Pin 11) emitter housing and the Controller circuit board ground (4 corner screws) to
prevent overheating or damage to the emitter. A 60 millivolt difference between the two grounds
will result in a 40°C error in the temperature sensor. This may likely result in damage to the
emitter AND VOID WARRANTY.
TO-8 EMITTER MECHANICAL DIMENSIONS
All dimensions are in inches except as noted.
Page 8