DPM TT 570 A Manuel d'utilisation - Page 14

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} DPM TT 570 A. DPM TT 570 A 16 pages. Micromanometer

TT Series LR Manual 
 
Page 12
 
Principle of Operation: 
Pitot Head 
 
Velocity Pressure* = Total Pressure – Static Pressure 
  
*
Calculated by the Micromanometer
Log Linear Rule for Traverse Points on 3 Diameters in a 
Circular Duct:
 
Alternative Measuring Points and Traverse Lines 
Relative to Side Lengths for Regular Ducts: 
 
DUCT TRAVERSE POINTS
SP
 
0.032D
0.135D
0.321D
0.679D
0.865D
0.968D
D
 
0.939D
0.765D
0.563D
0.437D
0.235D
0.061D
 
The nose of the Pitot Tube should face 
directly into the airstream thus the Total 
Pressure flows down the inner tube which 
TP
is connected to the Signal In port. 
The static holes are positioned around the 
side of the Pitot Tube and lead into an 
 
outer tube.  This is connected to the black 
tubing which in turn is connected to the 
Reference port. 
Ideally traverse points should be at least six 
duct diameters away from any bend or 
obstruction in the system.   
The Pitot Tube should be inserted at right 
angles to the walls of the ducts and 
measurements are taken in the positions 
shown in the diagrams (left). 
The directional pointer can be used to 
ensure that the Pitot Tube head is parallel 
to the duct walls.