Dr. Mach MACH 120 Series Instructions de montage, mode d'emploi - Page 15

Parcourez en ligne ou téléchargez le pdf Instructions de montage, mode d'emploi pour {nom_de_la_catégorie} Dr. Mach MACH 120 Series. Dr. Mach MACH 120 Series 19 pages. Ceiling, wall, new-type stand model
Également pour Dr. Mach MACH 120 Series : Manuel d'utilisation (19 pages)

Dr. Mach MACH 120 Series Instructions de montage, mode d'emploi
Wall attachment Mach 120/130
Mach LED 120/130
Springs are wearing parts, thus the spring force can decrease in time and must be readjusted.
Note: Set the spring force in such a way, that the spring arm with the lamp head holds its position
at every set height.
Dismount the lamp head in reverse order of mounting (see chapter 5.2).
Check the thickness of the segment. It must be at least 1,5mm. In case the thickness of the segment
is less than 1,5mm, it has to be changed.
Grease the segment and the coupling journal of the lamp.
Mount the lamp to the spring arm (see chapter 5.2).
59860002

7.2 Adjusting the spring force

B
A
B

7.3 Maintenance of the segment

Edition 11
In case the spring arm with the lamp moves
down on its own, proceed as follows:
Remove lateral cover A. The adjusting screw (B)
for the spring force is visible now.
Push the lamp and spring arm upwards as much
as possible, so the spring is without charge.
Insert a slotted screwdriver into the screw B.
Turn
the
adjusting
(anticlockwise), until the lamp holds its position
at every set height.
In case the spring arm with the lamp moves
up on its own, proceed as follows:
Remove lateral cover A. The adjusting screw (B)
for the spring force is visible now.
Push the lamp and spring arm upwards as much
as possible, so the spring is without charge.
Insert a slotted screwdriver into the screw B.
Turn
the
adjusting
(clockwise), until the lamp holds its position at
every set height.
25.25.2011 / Bak Page 15/19