8.4.1 alarm recovery
Probe alarms P1" and "P2" start some seconds after the fault in the related probe; they automatically
stop some seconds after the probe restarts normal operation. Check connections before replacing the
probe. Temperature alarms "HA" and "LA" automatically stop as soon as the temperature returns
to normal values.
Alarms "EA" and "CA" (with iF=bL) recover as soon as the digital input is disabled.
8.5 connection diagram
8.5.1XR01-02CX – 1 X 8A – 12VAC/DC
8.5.2 XR01-02CX – 20A OR 8A -- 110VAC OR 230VAC
NOTE: The compressor relay is 20(8)A or 8(3)A depending on the model.
NOTE: 120Vac connect to 6-7