5. Function Mode
The function mode stores parameters to control the weighing module.
The parameters are stored even without power supplied.
5.1.
Summary
............ Calibration function
............ Linearity adjustment function
............ Basics function
............ Hold function
............ Flow rate function
............ Control I/O function
............ Current loop output function
............ Analog 4-20mA output function
When only a numerical value is displayed, the decimal point flashes to distinguish
that from a weighing value.
5.2.
Operations and Types
5.2.1.
Select Modes Under Function Mode
..... Proceeds to function mode from the weighing mode.
+
........... Selects a type of select mode (3 upper figures).
....... Enters a selected mode.
........... Selects an item under the selected mode (2 lower figures).
....... Enters the item.
....... Stores parameters and returns to weighing mode.
5.2.2.
Change Value
........... Moves selected digit.
........... Changes numerical value.
....... Activates (store) value and returns to select mode.
....... Deactivates value and returns to select mode.
5.3.
Function Table
※1 Decimal point depends on
.
※2 When "span calibration using mass (
)" is performed by the input voltage from the
load cell in the calibration mode, "input voltage at zero (
)", "span calibration input
voltage (
)" and "weight against span calibration input voltage(
※3 If pressing the
key while displaying
or
, the current weighing value
can be monitored. Press the
key again to return function mode.
5.3.1.
Calibration Function (
)
Item & Function
Descriptions, Range & Factory settings
0 : No used
1: g
2 : kg
3: t
4: N
Unit
Decimal point (D.P.) position.
Decimal point position
0 : 0
1 : 0.0
2 : 0.00
3 : 0.000
Minimum division of value. (A scale / digit)
Minimum division
1 : 1
2 : 2
3 : 5
4 : 10
5 : 20
Measurement can be displayed up to +8 digits (8 scales) from
Weighing capacity
capacity. ※1
1 to 70000 to 99999
The range that the
key (zero key) works. Deviation from the
Zero range
calibrated zero point [%].
0 to 2 to 100
Used with
for zero tracking.
Zero tracking time
Scale: 0.1 sec.
Used with
for zero tracking.
Zero tracking band
Scale: 0.1 digit
Used with
for stability detection.
Stability detection time
Scale: 0.1 sec.
0.0 to 1.0 to 9.9
Used with
for stability detection.
Stability detection width
Scale: 1 digit
0 to 2 to 100
Tare and zero operation when unstable
Tare and zero at unstable
0 : Disables both functions.
1 : Enables both functions.
weight value
Tare when the gross weight is negative.
Tare when the gross
0 : Disables tare.
1 : Enables tare.
weight is negative
Output when out of
0 : Disables output.
1 : Enables output.
range and unstable
To judge when the negative gross weight is exceeded.
Exceeding negative
1 : Gross weight < -99999
3 : Gross weight < -19 digits
gross weight
2 : Gross weight < -Capacity
To judge when the negative net weight is exceeded.
Exceeding negative net
1 : Net weight < -99999
2 : Net weight < -Capacity
weight
Select whether or not to clear the zero value.
Clear the zero value
0 : Disable
1 : Enable
Select whether or not to perform zero setting when power is
Zero setting when power
turned on.
is turned on
0 : Disable
1 : Enable
-5-
Input voltage from a load cell at zero. ※2
Input voltage at zero
Scale: 0.0001 mV/V
Input voltage from a load cell at span. ※2
Span input voltage
Scale: 0.0001 mV/V
The calibration weight value corresponding to the input voltage at
Weight against span
. ※1 ※2
input voltage
Gravity acceleration of the place where the scale is calibrated.
Gravity acceleration of
2
Scale: 0.0001 m/s
the calibration place
Gravity acceleration of the place where the scale is being used.
Gravi t y accel e rati o n of use
2
Scale: 0.0001 m/s
pl a ce
Suppression of the hold
0 : Permission.
1 : Prohibition.
function
Reserved internally
5.3.2.
Linearity Adjustment Function (
)
Item & Function
Descriptions, Range & Factory settings
Number of adjustment point. If 0, 1 or 2 is selected, this linearization
Number of input points
is not used.
Input voltage for linear-zero input.
Linear-zero
Scale: 0.0001 mV/V
The setting value of weights for linear 1 input. ※1
Setting value for linear 1
The span voltage between linear-zero and linear 1 input.
Span at linear 1
Scale: 0.0001 mV/V
The setting value of weights for linear 2 input. ※1
Setting value for linear 2
The span voltage between linear-zero and linear 2 input.
Span at linear 2
Scale: 0.0001 mV/V
The setting value of weights for linear 3 input. ※1
Setting value for linear 3
The span voltage between linear-zero and linear 3 input.
Span at linear 3
Scale: 0.0001 mV/V
The setting value of weights for linear 4 input. ※1
Setting value for linear 4
The span voltage between linear-zero and linear 4 input.
Span at linear 4
Scale: 0.0001 mV/V
)" change.
5.3.3.
Basics Function (
)
Item & Function
Descriptions, Range & Factory settings
Each digit corresponds to to a key switch. Only available in weighing
mode.
4 fig. 3 fig. 2 fig. 1 fig.
Key switch disable
5: kN
0 : None
7: Zero clear
4 : 0.0000
1 : Manual print command 8: Reserved internally
2 : Hold
9: Reserved internally
6 : 50
3 : Operation switch 1
10: Reserved internally
4 : Operation switch 2
11: Reserved internally
key function
5 : Display exchange
12: mV/V monitor
6 : Tare clear
13: Digital filter 2
14: Display output data selected in
15: Display output data selected in
0.0 to 5.0
1 : 20 times/sec.
2 : 10 times/sec.
Display refresh rate
0 : None
0.0 to 9.9
1 : Zero tracking in progress
2 : Alarm
3 : Display operation switch status as on or off
display
4 : Near-zero
5 : HI output
6 : OK output
7 : LO output
Selects a cutoff frequency. ※3
0 : None
6 : 20.0 Hz
1 : 100.0 Hz
7 : 14.0 Hz
2 : 70.0 Hz
8 : 10.0 Hz
Digital filter 1
3 : 56.0 Hz
9 : 7.0 Hz
4 : 40.0 Hz
10 : 5.6 Hz
5 : 28.0 Hz
11 : 4.0 Hz
Selects a cutoff frequency.
0 : None
6 : 20.0 Hz
1 : 100.0 Hz
7 : 14.0 Hz
2 : 70.0 Hz
8 : 10.0 Hz
Digital Filter 2
3 : 56.0 Hz
9 : 7.0 Hz
4 : 40.0 Hz
10 : 5.6 Hz
5 : 28.0 Hz
11 : 4.0 Hz
1 : Normal hold
2 : Peak hold
Hold function
-6-
※1
-7.0000 to 0.0000 to 7.0000
Near-zero
0.0100 to 3.2000 to 9.9999
Comparison mass at
1 : Gross weight
near-zero
※1
1 to 32000 to 99999
Upper limit value
※1
Lower limit value
9.7500 to 9.8000 to 9.8500
Comparison mass of
1 : Gross weight
upper and lower limit
9.7500 to 9.8000 to 9.8500
5.3.4.
Hold Function (
Item & Function
Time to calculate the average. 0.00 is not averaged.
Average time
Scale: 0.01 sec.
Waiting time to commence a holding or averaging.
Start wait time
Scale: 0.01 sec.
The condition to commence a holding or averaging.
Condition of
0 : Not used
automatic start
1 : Above the near-zero, and stable
0 to 5
Release when control input of the hold terminal is falling.
Release using control
0 : Do not release
input
-7.0000 to 0.0000 to 7.0000
Release after a set amount of time has passed. 0.00 is not averaged.
Release time
Scale: 0.01 sec.
0 to 99999
Release when fluctuation from the holding value exceeds a set value.
Release using
※1
0.0000 to 9.9999
fluctuation range
0 : Continue
Release when the weighing value is in the near-zero range.
0 to 99999
Release at near-zero
0 : Do not release
0.0000 to 9.9999
5.3.5.
Flow Rate Function (
0 to 99999
Item & Function
0.0000 to 9.9999
Filter of flow rate 1
0 to 99999
Filter of flow rate 2
0.0000 to 9.9999
Damping time of flow rate 1
Damping time of flow rate 2
5.3.6.
Control I/O Function (
0 : Permission
1 : Prohibition
Item & Function
0 0 0 0 to 1 1 1 1
IN 1
0 : Not used
1 to 6 : Reserved internally 23 : Tare clear
IN 2
7 : Zero
8 : Tare
IN 3
9 : Hold
10 :
/ Net exchange 27 : Initialize flow rate 1
Gross
IN 4
11 : Diagnose
12 : Print command
IN 5
13 to 21 : Reserved internally 30 : Specify flow rate in
IN 6
3 : 5 times/sec.
OUT 1
0 : Not used
1 to 8 : Reserved internally
OUT 2
9 : Stability
10 : Over capacity
OUT 3
11 : Net display
12 : During tare
OUT 4
13 : Hold
OUT 5
14 : Hold busy
15 : HI output
OUT 6
16 : OK output
12 : 2.8 Hz
17 : LO output
13 : 2.0 Hz
OUT 7
18 : Near-zero
14 : 1.4 Hz
19 to 29 : Reserved internally
OUT 8
15 : 1.0 Hz
16 : 0.7 Hz
OUT 1
OUT 2
OUT 3
12 : 2.8 Hz
18 : 0.40 Hz
OUT 4
13 : 2.0 Hz
19 : 0.28 Hz
1 : Inverting output
OUT 5
14 : 1.4 Hz
20 : 0.20 Hz
OUT 6
15 : 1.0 Hz
21 : 0.14 Hz
16 : 0.7 Hz
22 : 0.10 Hz
OUT 7
17 : 0.56 Hz
23 : 0.07 Hz
OUT 8
3 : Averaging hold
5.3.7.
Current loop Output Function (
-99999 to 10 to 99999
Item & Function
2 : Net weight
Serial data
Communication mode
-99999 to 10 to 99999
Baud rate
-99999 to -10 to 99999
2 : Net weight
5.3.8.
Analog 4-20mA Output Function (
Item & Function
)
Descriptions, Range & Factory settings
0.00 to 9.99
Output data
0.00 to 9.99
2 : Above the near-zero
1 : Release
Mass/flow rate
at 4mA output
0.00 to 9.99
0 to 99999
Mass/flow rate
1 : Release
at 20mA output
)
Flow rate unit
Descriptions, Range & Factory settings
1 :Digital filter 1
Flow rate setting
12 :Digital filter 2
magnification
Suppress shaking of flow rate.
The higher value setting, the less shaking.
Scale: 1 sec.
1 to 5 to 1000
)
Output data
Descriptions, Range & Factory settings
0 to 7 to 30
22 : Zero clear
0 to 8 to 30
24 : Operation same as a F key
25 : Prohibit update of flow rate 1
0 to 30
26 : Prohibit update of flow rate 2
Mass/flow rate
at 4mA output
0 to 30
28 : Initialize flow rate 2
29 : Specify flow rate in
0 to 30
Select flow rate 1 or 2
0 to 30
Mass/flow rate
at 20mA output
0 to 18 to 36
0 to 9 to 36
30 : In weighing (ON)
0 to 36
Flow rate unit
31 : In weighing (1 Hz)
32 : In weighing (50 Hz)
0 to 36
33 : Alarm
34 : Output operation switch is
0 to 36
Flow rate setting
on or off
magnification
0 to 36
35 : Approximate flow rate value
of flow rate 1
0 to 36
36 : Approximate flow rate value
of flow rate 2
0 to 36
2 : Non inverting output
-7-
)
Descriptions, Range & Factory settings
1 : Displayed weighing value 3 : Net 5 : Gross / Net / Tare
2 : Gross
4 : Tare
1 : Stream
3 : Manual print
2 : Automatic print
1 : 600 bps
2 : 2400 bps
)
Descriptions, Range & Factory settings
1 : Displayed weighing value (Digital filter 1)
2 : Gross (Digital filter 1)
3 : Net (Digital filter 1)
4 : Displayed weighing value (Digital filter 2)
5 : Gross (Digital filter 2)
6 : Net (Digital filter 2)
7 : Flow rate 1
8 : Flow rate 2
9 : Flow rate 1 or Flow rate 2 (Select in control input)
Select mass/flow rate by setting output data (
)
Decimal point position linkage:
Mass
Flow rate
(setting magnification of flow rate) +
-99999 to 0 to 99999
Select mass/flow rate by set output data (
)
Decimal point position linkage:
Mass
Flow rate
(setting magnification of flow rate) +
-99999 to 70000 to 99999
1 : Seconds
2 : Minutes
3 : Hours
1 :
1
2 :
10
3 :
100
(times)
4 :
1000
5 : 10000
1 : Displayed weighing value (Digital filter 1)
2 : Gross (Digital filter 1)
3 : Net (Digital filter 1)
4 : Displayed weighing value (Digital filter 2)
5 : Gross (Digital filter 2)
6 : Net (Digital filter 2)
7 : Flow rate 1
8 : Flow rate 2
9 : Flow rate 1 or Flow rate 2 (assign to input terminals)
Select mass/flow rate by set output data (
)
Decimal point position linkage:
Mass
Flow rate
(setting magnification of flow rate) +
-99999 to 0 to 99999
Select mass/flow rate by set output data (
)
Decimal point position linkage:
Mass
Flow rate
(setting magnification of flow rate) +
-99999 to 70000 to 99999
1 : Seconds
2 : Minutes
3 : Hours
1 :
1
2 :
10
3 :
100
4 :
1000
(times)
5 : 10000
-8-