Technical Data

Dimensions
Weight
Temperature measurement range Ear/Forehead mode: 34 to 42.2°C(93.2 to 108°F), Surface mode: -22 to +80°C(-7.6 to 176°F),
Operating temperature range
Storage temperature range
Accuracy
Memory
Battery
・This thermometer converts the forehead/ear temperature and displays the "oral equivalent".
・Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
NOTE: Specifications are subject to change for improvement without prior notice.
125 mm × 37 mm× 22 mm
Approx. 41g, excluding the battery.
10 to 40°C (50 to 104°F) ,15 to 85%RH
Device should be stored at a room temperature between -20 to +50°C and humidity 85% or less.
Transportation temp. shall be less than 70°C ,RH≦95%, Atmospheric pressure:800 to 1013hPa.
Ear/Forehead mode: ±0.2°C (0.4°F) within 35 to 42°C (95 to 107.6°F) (Ambient Temp: 15 to 35°C)
,±0.3°C (0.5°F) for other range.
Surface mode: ±0.3°C (0.5°F) within 22 to 42.2°C (71.6 to 108°F)
others ±4% or 2°C (4°F) whichever is greater.
25 memory slots and last memory.
CR2032×1
English 13