Troubleshooting

1. The AlcoSense AccessPoint cannot power on
Reason: Power supply is not connected correctly
Solution: Check that the red cable voltage is 12V.
If problem still persists, please contact the manufacturer.
2. Red Status l amp
Reason: Breath sampling is not strong enough for a breath analysis (Flow error )
Solution: Wait until the Status lamp is green, then blow steadily into the sampling hole.
3. Al l 3 l amps (Status, Pass and Fail l amps) keep bl inking
Reason: White cable is connected to the door sensor or door is not closed properly
Solution: Check the door sensor and whether the door is opened or closed
4. Analyzing and showing resul t is ver y slow, or no resul t provided after breath sample:
Reason: The AccessPoint sensor module needs calibration or replacing. If the unit is not providing a
result, the flow sensor in the sensor module may be damaged.
Solution: Please contact the manufacturer.

Specifications

Indication of B.A.C.
Accur acy
Sensor
Mouthpiece
Power suppl y
Power r equir ements Minimum 12V DC 1amp (Amp requirement varies with connected components)
War m up time
Response time
Cal ibr ation inter val
Ambient conditions
Dimensions
Weight
Test result indicated by LED display
PASS : Indicate Green Lamp
FAIL : Indicate Red Lamp
(Level for PASS can be adjusted)
±10% at 0.500 g/L
Replaceable fuel cell sensor
Not required
DC12V 1.5A
3 seconds ~ 4 minutes
Within 3 seconds at 0.000 g/L
Within 10 seconds at 1.000 g/L
Recommended 12 months or 1,000 tests
Storage: 0 to 50
/ Operation: 5 to 40
206 (height) × 108 (width) × 37 (thickness) mm
674g
9
Version Number
Date Updated
V1.1
2.12.2015