Andatech PRODIGY 3 Manuel de l'utilisateur - Page 6
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Andatech PRODIGY 3. Andatech PRODIGY 3 14 pages. Industrial breathalyser
CONDITIONS FOR TESTING
MEASUREMENTS & CONDITIONS FOR TEST
• Use a clean mouthpiece for each test.
• The optimum temperature range for operating the unit is -10°C to 50°C. Use within this
temperature range will limit condensation from a subject's breath and ensure an accurate
reading is taken.
• Do not allow samples containing smoke,tobacco or other chemical products to enter the unit, as
damage to the fuel cell sensor can occur.
• It is recommended that the operator observe the subject for approximately 20 minutes before
administering the breath test. This is to ensure the subject does not consume anything.
Following this procedure will ensure the subject's saliva is clear of alcohol so that it does not
alter the outcome of the breath sample test.
• There must be an interval of at least 15 minutes after alcohol consumption as alcohol residue
in the mouth may give a false reading. Such residue may also be left by aromatic drinks,
mouthwash, medicines and drops, as well as by belching and vomiting. Mouth rinsing with water
or non-alcoholic drinks does not substitute for an interval.
• A person's blood alcohol concentration (BAC) can continue to rise for up to 2 hours after
drinking alcohol. If a result is close to the designated BAC, it is recommended to wait and retest
after 20 minutes.
• It can take 10 hours or more for a person's BAC to return to zero after a high BAC has been
reached. A further test should be carried out later in the day or the following morning.
LIABILITY FOR PROPER FUNCTION OR DAMAGE
The liability for proper functioning of the instrument is irrevocably transferred to the owner
or operator in the event the instrument is serviced or repaired by personnel not employed
or authorised by Andatech. In such an event, the warranty will be void.
OPERATING CONDITIONS
Working temperature of the Andatech® Prodigy 3: -10 to 50°C
Storage temperature range of the Andatech® Prodigy 3: -30 to 70°C, 15 to 90%RH
Note: Inappropriate storage of the device will shorten the lifespan of the fuel cell sensor.
6