Section 4 - Wiring
4.1 SIGNAL PROBE WIRING
Always route signal cables separate from AC power and AC control circuits. Signal cables must be shielded
with the bare shield drain wire grounded on one end and floating on the other end.
NOTE: If it becomes necessary to splice cable connections, be sure to properly solder or join wire
connections. Loose connections have the ability to generate false counts.
DIAGRAM 4.2 WIRING DIAGRAM
FEMALE SOCKET (1)
BLACK WIRE
MALE PIN (3)
PIN/SOCKET ORIENTATION
FEMALE SOCKET (1)
BLACK WIRE
MALE PIN (2)
RED WIRE
PIN/SOCKET ORIENTATION
ANDERSON INSTRUMENT CO., INC.
WHITE WIRE
OUTPUT WIRING
BLACK
(
SIGNAL OUTPUT
WHITE
NO POLARITY
SHIELD
ANDERSON INSTRUMENT CO., INC.
MALE PIN (3)
WHITE WIRE
INPUT/OUTPUT WIRING
RED (SUPPLY INPUT+8-24VDC)
BLACK (COMMON)
WHITE (SIGNAL OUTPUT)
SHIELD
PART NUMBER HT750000
HT750000 LENGTH - APPROX. 12''
WHEN THREADED INTO HM SERIES FLOWMETER
HT750000 PICK-UP PROBE SPECIFICATIONS
SIGNAL OUT
OPERATIONAL FREQUENCY RANGE
OUTPUT VOLTAGE
V
)
RED + BLK COM WH SIG
PART NUMBER HTS1000
HTS1000 LENGTH - APPROX. 12''
WHEN THREADED INTO HM SERIES FLOWMETER
SIGNAL OUT
EXCITATION VOLTAGE
OPERATIONAL FREQUENCY RANGE
DUTY CYCLE
V
40%
60%
40%
TURBINE FLOWMETER
.18 - 2.6 V.P .P .
Hz
TURBINE FLOWMETER
SQAURE WAVE
8-24 VDC, 0.8 Watt max
0-1000Hz
40 HIGH - 60 LOW
~
SUPPLY VOLTAGE MINUS 0.7V
~
0.5V
T
HMP SERIES
SINE WAVE
0-1000Hz
HMP SERIES