ANDERSON-NEGELE MPP Manuel d'installation et de démarrage
Parcourez en ligne ou téléchargez le pdf Manuel d'installation et de démarrage pour {nom_de_la_catégorie} ANDERSON-NEGELE MPP. ANDERSON-NEGELE MPP 4 pages. Modular pressure transmitter
Anderson Instrument Co., Inc.
156 Auriesville Rd., Fultonville, NY 12072
Phone: 518-922-5315
Fax: 518-922-8997
www.anderson-negele.com
SENSOR WIRING
To facilitate electrical connections the MPP transmitter will be provided with either a 5 pin M12 quick disconnect
receptacle, a M16 thread cable gland, or a ½" NPTF threaded adaptor. Shielded cable is recommended. See manual for
additional detail.
Field wireable connectors or molded cordsets are available as accessories from Anderson Instrument.
FIELD WIREABLE CONNECTOR ASSEMBLY
1. Insert cable through Pressing Screw, Compression
Ring, Seal Grommet, and Sleeve as shown below.
2. Strip back 1-1/4" of outer sheathing, cut off any excess
wires, shield and ground. Strip off 1/4" insulation from
remaining two wires. It is not necessary or recommended
Shown with
wire cover plate
removed
Swtiches and display only on
HART version
MOLDED CORD SETS
Anderson Cord Set
READ
THIS
FIRST
PRODUCT DESCRIPTION
The model MPP modular pressure transmitter has been specifically designed for
placement in sanitary process applications in the Life Sciences industry where
accurate and repeatable pressure measurements in difficult (vibration, wide
temperature swings, humidity) environments are required. The modular design
allows for field configuration to best suit application needs and allows economical
component replacement. State of the art performance yields class leading
accuracy and stability while minimizing process and ambient temperature effects. A
menu driven interface with an internal diagnostic display provides user adjustability
of pressure range and a host of other control characteristics without tools or
pressure standards.
3. Orient Connector end so that
Retaining Ring
Connector End
Pin 2 - Black
(Loop -)
Loop+(red) wire Pin 1
Loop- (black) wire Pin 2
Shield (clear or bare) wire
Note: Green and White not used
Installation and Startup Guide
MPP Modular Pressure
Version 2.1
- ORDERED AS ACCESSORY
center pin connecting screw is
horizontal facing right (see
detail).
Sleeve
Center screw pin
orientation reference
Pin 1 - Red
(Loop +)
DETAIL
Transmitter
Document 1184
4. Wire LOOP+ (red) wire to top-right
terminal, and LOOP- (black) wire to
top-left terminal. No connection is
made to the center and bottom
terminals.
5. Screw on the Sleeve. Hand-tighten only.
Compression
Ring
Pressing Screw
Seal
-2 included choose one to accommodate cable OD
Grommet
6. Press the Seal Grommet into the Sleeve and
hand-tighten the Pressing Screw against the
compression ring.
7. Use a wrench to tighten the Pressing Screw
another 3/4 turn. Do not over-tighten!
Typical Cord Set by others
Loop+(brown) wire Pin 1
Loop- (white) wire Pin 2
Shield (bare) wire