Atkins AquaTuff 353 Manuel de l'utilisateur - Page 4

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Atkins AquaTuff 353. Atkins AquaTuff 353 6 pages. Digital thermocouple thermometer warranty and user guide

Atkins AquaTuff 353 Manuel de l'utilisateur
AquaTuff™ 353 THERMOMETER
ITS MODE - Continued:
2) When ITS MODE is active, the display will show "- -"
while the temperature is changing (stabilization
seek mode). The final stabilized temperature reading
will flash on the display when reached. The ITS
temperature reading will remain flashing on the
display until the probe is removed or temperature
changes by approximately 5°F(3°C). When the
temperature changes, the instrument will return to
ITS stabilization seek mode and "- -" will be
displayed until the next stabilized temperature is
reached and that temperature reading will flash on
the display.
3) The user has the option of storing the ITS
temperature reading in memory. To store ITS
temperature reading, press
reading. The display will show reading number
"r001-r250" and store reading in memory. Up to 250
readings can be stored in memory.
4) To return to STANDARD TEMPERATURE MODE,
turn unit OFF and back ON or repeatedly press
key until both ITS and MEM icons are
turned off at top of display.
MEMORY MODE:
1) To view temperature readings stored in memory, turn
instrument ON and press
MEMORY MODE. When active, MEM icon will
appear at top of LCD. (NOTE: WITH NO
TEMPERATURE DATA IN MEMORY, INSTRUMENT
WILL NOT ENTER MEMORY MODE.)
2) To view temperature data, press
forward or
key to scroll backward through
temperature data. The display will show reading
number "r001-r250" followed by the temperature
recorded. (NOTE: TO ACCESS FAST SCROLL
MODE, PRESS AND HOLD
KEY AND DISPLAY WILL QUICKLY SEQUENCE
THROUGH TEMPERATURE READINGS.)
3) The temperature data is a cyclical data system.
When scrolling through data, the temperature
readings will cycle through all readings either forward
or backward and then repeat readings when either
the first or last reading in memory is reached.
4) To clear temperature readings in memory, press and
hold both the
and
time until "CLr" is shown on display.
5) To return to STANDARD TEMPERATURE MODE,
turn unit OFF and back ON or repeatedly press
key until both ITS and MEM icons are turned
off at top of LCD display.
6
key to store
key to select
key to scroll
KEY OR
keys at the same
AquaTuff™ 353 THERMOMETER
MEMORY MODE - Continued:
USER TIP
When temperature readings are cleared
from memory, there is no facility to
restore deleted temperature data.
Ensure that temperature data is not needed
before clearing memory contents.
351 / 352 / 353 BATTERY REPLACEMENT
BATTERY REPLACEMENT:
1) Turn battery door knob counter-clockwise until
rotation stops. The arrow on battery door knob will
point towards unlock
icon.
2) Pull battery door knob to remove battery door
assembly. Remove batteries.
3) Referring to battery polarity drawing on rear label,
install (2) AAA alkaline batteries as shown.
4) To install battery door assembly, ensure that battery
door knob is turned fully counter-clockwise and
arrow on battery door is pointed towards unlock
icon. Push upward on battery door assembly until
battery door knob is seated against housing. Rotate
battery door knob clockwise until rotation stops.
The arrow on battery door knob will point towards
lock
icon.
TURN COUNTER-CLOCKWISE TO UNLOCK
TURN CLOCKWISE TO LOCK
7