2.3
Choosing the infusion line
Different types of infusion lines may be used. Their choice and handling with regard to air drainage are
determined according to the Droper's mode of use.
1. Infusion line with drip chamber positioned in the middle of the infusion line (not with drip
chamber connected directly to the spike).
b. Infusions in hostile environments with drip chambers capable of being positioned lying
down, or upright with liquid output in a upwards direction during transfer under
infusion: in this case there is a risk that air bubbles originating from the free air
enclosed in the drip chamber are infused. The operator must therefore use the reverse
purge air drainage technique expelling all air including that contained in the drip
chamber. The drip chamber will no longer be used as a reference to estimate the flow
of solution infused. This flow must be estimated or checked using a suitable flow
regulator.
a. Residential infusion (Home care, Hospital, Field
hospital, ...) with drip chamber positioned upright,
liquid output in a downwards direction (classic
example): as for gravitational infusions, the drip
chamber is filled with the perfusion fluid until a
certain height is reached, with air being
positioned at the top of the drip chamber. The
flow will be estimated by counting the number of
drops in the first minute of infusion.