Droper Field 1000 Manuel d'utilisation - Page 9

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Droper Field 1000. Droper Field 1000 15 pages. Mechanical pressure infuser
Également pour Droper Field 1000 : Manuel d'utilisation (16 pages)

Infusion line with « Air Stop » filter
Essential precaution : Perform the classic purge
= the line is purged of any air and the drip chamber is half filled.
1. Close the infusion line roller clamp
2. Insert the spike into the infusion bag like for a classic infusion
3. Place the bag in the horizontally arranged Droper Field
4. Close the Droper Field
5. Fill the drip chamber halfway by pinching it between thumb and
index finger
6. Make sure that the drip chamber does not position horizontally but is tilted
downwards to allow permanent contact of the Air Stop filter with the liquid. This
can be facilitated by lines that have an extension between the spike and the drip
chamber (see aside drawing)
7. Open the infusion line roller clamp
8. Purge the line of any air like during classic infusion procedure
9.
Close the infusion line roller clamp
10. Connect the infusion line to the patient and re-
open the roller clamp
Flow regulation : through the drip chamber
4
7
9
Instructions for use - Version 2
1
2
5
8
10
3
6
9/15