Droplet SP2 Manuel - Page 7
Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} Droplet SP2. Droplet SP2 13 pages. Beam-scan camera and software
Manual, SP2 Beam Scan Camera and Software
Figures 1 -3 show the data obtained for a YAG laser beam when set properly. These plots show
the view of the beam effectively end-on in the control panel, and the beam profile in the plot
and surface views. The measurement will appear to have considerable jitter in the plot view.
This is caused by the short exposure time. If the beam profile appears basically Gaussian,
(normal or even distribution) but the top of the view is flat in the plot or profile views, this is
caused by measurement saturation. The exposure time should be reduced to change this.
Figure 1: SP2 Laser Mode in the TEM
Mode, as Viewed in the vBeam Control Panel
oo
7
DOC-0175 Rev D
© 2017 DROPLET MEASUREMENT TECHNOLOGIES