15.
Check the forward stoke of the paddles.
a.
Rest the squared-off end of the depth gauge against the
face of the orange gasket on Clamp Bar B just in front
of paddle #5.
b. Using either paddle #1 or #6 in your opposite hand,
rotate the paddles forward so that they extend to their
forward position.
c. Looking straight down on paddle #5, measure the
distance that the bottom edge of this paddle extends
beyond the orange gasket. This distance should be
4/32" - 5/32". If this distance is not correct contact
ANKOM Technology for instructions on how to make
an adjustment.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091