Drucker 642E Manuel de l'opérateur - Page 4
Parcourez en ligne ou téléchargez le pdf Manuel de l'opérateur pour {nom_de_la_catégorie} Drucker 642E. Drucker 642E 6 pages. Laboratory centrifuge
well as premature failure of the centrifuge. Choose an area which will allow
unencumbered air flow.
d)
The centrifuge is designed to secure to the operating surface by four suction
feet. No adjustment is necessary for leveling the centrifuge, however, the
surface should be flat and level.
e)
Be sure the outlet is always within reach as the line cord is the means of
emergency disconnection!
Initial Setup Procedure:
If any problems are found during the initial setup procedure, refer to the
troubleshooting section on page 10. For further assistance, contact
Drucker Diagnostics at
814-342-6205 or
1.
Plug the centrifuge in to an approved electrical outlet. For electrical safety, the unit
must always be properly grounded.
2.
For safety purposes, the locking system is always activated. To deactivate the
system, (in order to insert or retrieve samples), press the 'OPEN / STOP' button on
the control panel. The 'UNLOCKED' indicator light should illuminate. If it does not,
refer to page 10 on troubleshooting. The lid will be unlocked for 15 seconds after
pushing the 'OPEN / STOP' button.
3.
Turn the latch counterclockwise and open the lid.
4.
Spin the rotor by hand; check for free and level rotation. If the rotor does not spin
freely, refer to page 10 on troubleshooting.
5.
Place the six test tube holders inside the rotor (as shown to the right), and verify
that they are seated properly.
6.
Close the lid. Rotate the lid knob clockwise to its complete
stop position.The 'LATCHED' indicator light should be
illuminated. If it is not, make sure that the lid is latched
properly. The centrifuge will not run unless the lid is latched
and that the 'LATCHED' light is on.
7.
Turn the centrifuge on by pushing the 'START' button.
8.
The 'RUNNING' indicator light will illuminate.
9.
The test tube holders will slide up into the horizontal position and the unit will
accelerate to full speed.
10. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If
there are any loud or unusual sounds, stop the centrifuge by pushing the 'OPEN /
STOP' button immediately and refer to page 10 on troubleshooting.
11. While the machine is running, try to turn the latch counterclockwise. Power may
be cut to the motor but you should be unable to fully turn the latch. If it is possible
to turn the latch and open the lid while the unit is running, contact Drucker
Diagnostics for assistance. Close and latch the lid.
12. Push the 'OPEN / STOP' button. The 'RUNNING' indicator light should
go out and the motor should slow to a stop.
13. The lid should remain locked until the rotor has nearly stopped. If
the machine unlocks prematurely, contact Drucker Diagnostics for
assistance.
Once the rotor has stopped, the interlock system will become
disengaged for sixty (60) seconds. The 'UNLOCKED' indicator light will
illuminate during this time.
14. To gain entry into the centrifuge after this period has ended, simply press
the 'OPEN / STOP' button. The lid will unlock for fifteen (15) additional
seconds.
After the centrifuge has passed this procedure it is ready for operation.
814-692-7661.
Page 4
glass fragments left in the tube holder after a tube breakage may adhere to
the next test tube inserted in that holder. When this tube is handled, these
fragments may puncture protective gloves and lacerate the operator's
fingers or hand. Remaining fragments may provide stress points on
subsequent tubes and result in additional breakage. If a tube breakage
occurs, carefully remove the tube holder. Properly dispose of the sample
and tube fragments and thoroughly clean both the inside and outside of the
tube holder. Insert a new tube cushion (if necessary) and replace the tube
holder in the rotor.
4.
Motor and Electrical Maintenance: The Model 642E uses a brushless
permanent split capacitor AC motor. It should not need servicing for the life
of the centrifuge. The electrical components are selected for high reliability
and should not need service.
5.
Tube Holder Replacement: It is recommended that the tube holders be
replaced after 24 months of use. Inspect tube holders regularly for cracks. If
cracks are discovered, replace immediately.
6.
Remove Accessories Before Moving: All tube holders, samples, and caps
must be removed from the rotor chamber before transporting or storing the
centrifuge to prevent damage and injury.
Cleaning and Disinfection:
To prolong the life of the centrifuge cleaning and disinfection is recommended every
six months, or whenever there is a spillage or tube breakage. Contaminants must
be removed immediately, or corrosion and premature degradation of components
can occur.
1.
Unplug the centrifuge before cleaning.
2.
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge
in water or other cleaning solutions as this will cause damage and void the
warranty.
3.
ONLY isopropyl alcohol, soap and water, or a 10% (5500 PPM) bleach
solution should be used for cleaning and disinfection of the centrifuge and
accessories.
4.
All surfaces must be dried immediately after cleaning and disinfecting.
5.
TBQ Germicidal products shall not be used, as they will cause damage to the
centrifuge and void the warranty.
6.
The use of fully/partially halogenated hydrocarbons, ketones, esters, ethers,
benzyls, ethyl benzenes, and all other chemicals not prescribed by the
manufacturer shall not be used as they may cause damage to the rotor
chamber, rotor, tube holders, accessories and centrifuge exterior and void the
warranty.
7.
It may be necessary to remove the rotor and clean the rotor chamber. Follow the
instructions on page 8 to remove and reinstall the rotor.
(cont.)
Page 9