Transducer Instruction Manual
Range:
Ultra
Type:
CLT
Revision:
V1.0
4.4
Output Signal Not Linear, Zero Shifts During Operation
[1]
Check that there is no dirt or foreign matter interfering with the transducer mounting.
[2]
Check that the shaft clamp and transducer mounting bolts are secure.
4.5
Transducer Strain Gauge Resistance
The transducer has semiconductor strain gauges in a full bridge configuration.
The gauge resistances, measured at room temperature with no load, are shown in the table below.
Measurements taken
between the following
wires:
GN/RD
YL/RD
GN/YL
GN/BU
YL/BU
RD/BU
_____________________________________________________________________________________________________________________
Absolute Measurement Ltd. (T/A CMC Controls)
t: (0) 1732 763278
| e:
Unit 19
| Silver Birches Business Park | Aston Road | Bromsgrove | B60 3EU
Corresponding
M12 Connector
Target Resistance
Pins:
Aluminium Beam
2 - 1
4 - 1
2 - 4
2 - 3
4 - 3
1 - 3
| w: absolutemeasurement.co.uk
Target Resistance Steel Beam
88 Ω
88 Ω
118 Ω
(Varies slightly based on load cell
104 Ω
(Varies slightly based on load cell
104 Ω
(Varies slightly based on load cell
133 Ω
88 Ω
88 Ω
118 Ω
112 Ω
temperature)
112 Ω
temperature)
141 Ω
temperature)
8