True Definition Scanner Wand
Cleaning and high level disinfection are two distinct procedures
required to reprocess the 3M
safe reuse.
Step 1: Wand Cleaning
Meets regulatory agencies' requirements
1. Use fresh gloves prior to unplugging and cleaning.
2. Unplug the wand using the boot (textured portion) of the connector and loop the cord to make it portable.
3. Rinse wand surface under running tap water (~22°C/72°F) for one minute without getting cable strain relief or USB connector wet.
4. Wipe wand surface for thirty seconds with a soft, non-abrasive wipe wetted with an enzymatic detergent solution.
5. Discard wipe into biohazard waste.
6. Rinse wand under running tap water for one minute without getting cable strain relief or USB connector wet.
7. Dry with clean, non-abrasive, lint-free wipe and allow to dry completely.
8. Inspect wand for any visible contamination. If needed, repeat cleaning process.
Step 2: Wand Disinfecting
Meets regulatory agencies' requirements
1. Immerse wand in an appropriate soaking container containing one of the high level disinfectants listed in the table below.
No CaviWipes.
2. Follow manufacturer's instructions for use and label information for disinfectant used. Note prescribed soak times.
3. After soaking, rinse wand under running water for one minute.
4. Dry with clean, non-abrasive, lint-free wipe and allow to dry completely.
5. Before use, reconnect the wand by plugging wand connector into USB port.
6. To keep wand clean prior to next use, a barrier sheath is recommended.
Step 3: Pre-Use Check
After disinfection and before using the 3M
visual check of the system. Look for cracks in the wand or cable, severe kinks in the cable, metal discoloration, corrosion
or damage of any kind. Check to make sure the wand and lens are clean.
Disinfectants for high level disinfection of the wand
Active Ingredient Minimum Concentration
0.55% Ortho-phthaldahyde (OPA)
2% Glutaraldehyde
* High level disinfection validations were conducted for the 3M True Definition Scanner Wand using Cidex OPA, Dentasept Ultra and Metricide Plus 30. If these are not available,
determine a compatible equivalent disinfectant based upon regulatory clearance (high level disinfectant), equivalent or higher active ingredient concentration and listed pH range.
Failure to adhere to the proper protocol listed in the steps above, may lead to improper disinfection and potential cross
!
contamination hazards.
True Definition Scanner wand for
True Definition Scanner, we recommend a dental professional perform a
Examples of Tradenames*
Cidex
OPA, Rapicide
OPA,
®
®
Metricide
OPA
Dentasept
Ultra, Metricide
Plus 30,
®
ProCide-D
Plus, Aldahol
HLD
®
®
Cleaning and Disinfection
Quick Reference Guide
3M
pH Range
Directions for Use
Follow disinfectant label instructions
7.2-7.8
for soak time and conditions
Follow disinfectant label instructions
6.5-10
for soak time and conditions
True Definition Scanner
Videos available
on 3M Dental
YouTube playlist