Electrical Safety and Compatibility

Basic EMC standard
Phenomenon
Operating mode
or test method
Mains terminal
CISPR 11:2015
N/A
disturbance voltage
Continuous
Radiated disturbance
CISPR 11:2015
operation mode
EN 61000-3-2:2014
Harmonic Current Emission
N/A
IEC 61000-3-2:2014
Voltage change,
EN 61000-3-3:2013
Voltage fluctuations and
N/A
IEC 61000-3-3:2013
Flicker Emission
Electrostatic Discharge
EN 61000-4-2:2009
Continuous
Immunity
IEC 61000-4-2:2008
operation mode
Radiated RF Electromagnetic
EN 61000-4-3:2006+A2:2010
Continuous
Field Immunity
IEC 61000-4-3:2010
operation mode
Immunity to Proximity
EN 61000-4-3:2006+A2:2010
Fields from RF wireless
N/A
IEC 61000-4-3:2010
Communications Equipment
Electrical Fast Transient
EN 61000-4-4:2012
N/A
/ Burst Immunity
IEC 61000-4-4:2012
EN 61000-4-5:2014
Surge Immunity
N/A
IEC 61000-4-5:2014
Immunity to
EN 61000-4-6:2014
Conducted Disturbances
N/A
IEC 61000-4-6:2013
Induced by RF fields
Power Frequency Magnetic
EN 61000-4-8:2010
Continuous
Field Immunity
IEC 61000-4-8:2009
operation mode
EN 61000-4-11:2004
Voltage dips
N/A
IEC 61000-4-11:2004
EN 61000-4-11:2004
Voltage interruptions
N/A
IEC 61000-4-11:2004
22
04
Appendix
Port tested
Test Voltage
Test level / Requirement
N/A
N/A
N/A
Enclosure
Battery
Group 1, Class A
N/A
N/A
N/A
N/A
N/A
N/A
± 8
kV/Contact
Enclosure
Battery
± 2, ± 4, ± 8,
± 15 kV/Air
3 V/m
Enclosure
Battery
80 MHz-2.7 GHz
80% AM at 1 kHz
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
30 A/m
Enclosure
Battery
50 Hz & 60 Hz
N/A
N/A
N/A
N/A
N/A
N/A
N/A
Symbols
Symbol
Description
Symbol
Description
Manufacturer
MR unsafe
Date of manufacture
Atmospheric pressure
and country of
manufacture
limitation
KR
(made in Korea)
Unique Device
Humidity limitation
Identifier
Model number
Temperature limit
Keep away from
Serial number
sunlight
Refer to instruction
Do not use if package is
manual
damaged and consult
/ booklet
instructions for use
04
Appendix
Symbol
Description
Symbol
Description
Ingress Protection
Caution
Waste electrical and
Type BF applied part
electronic equipment
Direct Current
Medical Device
Prescription Only
Professional Use Only
Do not re-use
Not available without a
doctor's prescription
Use by It can be used
up to the date stated
Keep dry
by the manufacturer
Authorized representative
CE Marking of
in the European
Conformity
Community
/ European Union