Auber Instruments AT200CHIM Manuel d'instruction - Page 2

Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Auber Instruments AT200CHIM. Auber Instruments AT200CHIM 2 pages. Thermometer for stove pipe, chimney, w/ remote alarm

AUBER INSTRUMENTS
6. Alarm setting (code 0001)
The meter has two programmable alarms that can be set to turn on the buzzer at a
specific temperature. The alarm is controlled by parameters AH1, AL1, AH2 and AL2.
It can be accessed by code 0001. AH1 and AH2 should be set to the temperature
that alarm turns on. AL1 and AL2 are set to the temperature Alarm turns off. When
AH1>AL1(or AH2>AL2), the alarms is set for hi g h limit alarm. When AH1<AL1(or
AH2<AL2), the alarm is set for low limi t alarm. For exampl e , if AH1=900, AL1=800(
default settings), when the temperature goes up to 900° F, the buzzer will be on;
when the temperature drops down to 800° F, the buzzer will be off. However, if
AH2=180, AL2=185, when the temperature drops down to 180° F, the buzzer will be
on; when the temperature goes up to 185° F, the buzzer will be off. User can press
the shift key (>) to temporari l y mute the buzzer sound. The alarm will buzz again if
the alarm set temperature i s reached again. To permanently deactivate the alarm,
set AH1=AL1 and AH2=AL2.
How to set alarm:
(1) Press set , enter code 0001, and set again to enter alarm setting
menu;
(2) Press ^ and v to select AH or AL;
(3) Press set to view the value of the parameter.
(4) Press >,^ and v to enter a new value.
(5) Press set to confirm;
(6) Press ^ or v to select the new parameter.
(7) To exit the menu, press set when End is displayed
The procedure is shown as the following flow chart in figure 4.
Oper ation Mode
Enter Code
SET
SET
0001
XXXX
SET
Figure 4. Flow chart of how to set the alarm
7. System configuration (code 0089)
Note: If you purchased the thermometer as a bundle (including the temperature probe),
there is no need to change any of these parameters. But if you want to recalibrate the
meter or reconfigure it, you need to read this section.
Table 1. System Configuration Parameters
Code
Description
inty
Input Sensor Type
Inty
psb
Input Offset
PSb
Digital Filter
FILT
FILT
C-F
C-F
Display Unit
Brit
brit
Brightness control
end
End
Exit
Note for each parameter:
1. Inty. Input sensor type. This parameter defines the sensor type that is used for
the thermometer. If you purchased the thermometer bundle, it is already set for the
probe included in the package. You don t need to change it.
2. PSb. Calibration offset. PSb is used to set an input offset to compensate the
error produced by the sensor. For example, if the meter displays 5 ºC when probe is
in ice/water mixture, set PSb= -5, will make the meter display 0 ºC.
3. FILt. Digital Filter. If measurement input fluctuates due to noise, a digital filter
can be used to smooth the input. FILt may be confi g ured in the range of 0 to 3.
Stronger filtering increases the stability of the readout display, but causes more
delay in response to the change in temperature. FILt=0 disables the fi l ter.
4. C-F. Temperature unit. Set to C i f you want to displ a y Celsius. Set to F for
Fahrenheit.
5. Brit. Brightness control. This parameter controls the LED display brightness.
(The default value i s 4 for the brightest display).
6. End. Exit the parameter setting mode.
2013.11
Parameter Display
Buzzer On Temp
SET
0480
AH1
Buzzer Off Temp
SET
0479
AL1
Buzzer On Temp
SET
0250
AH2
Buzzer Off Temp
SET
0250
AL2
END
Setting Range
Initial
-320~2400 ° F
K
-100~100(deg)
0
0
0~3
° C ° F
° F
1-4
4
The flow chart in figure 5 shows how the system configuration parameters
can be accessed.
(1) Press set , enter code 0089, and set again to enter system
confi g uration menu;
(2) Press ^ or v to select the parameter to be changed.
(3) Press set to view the value of the parameter.
(4) Press >,^ and v to enter a new value.
(5) Press set to confirm;
(6) Press ^ or v to select the new parameter.
(7) To exit the menu, press set when End is displayed.
SET
SET
SET
SET
SET
Figure 5. Flow chart of how to set the parameters
SET
8. Placing the meter
The operating environment temperature for the meter is from -20-50C (or, 0-100F).
The meter should be placed away from high heat to protect the plasti c housing and
electronics. Two mounting opti o ns are provided.
1)
A pair of Velcro fastener. The Velcro has a pressure sensi t ive adhesive
backing. You can remove the protecting film from the hook piece and stick it to the
back of the meter (see figure 6). Then remove the protecting film from the loop piece
and instal l it onto the wall. Please note that the pressure sensitive adhesive on Vel c ro
is industrial grade with strong holding force. It needs to be stick on solid surface. Don t
put it on a drywall because it may peel off the paint if you decided to remove it later
2)
A stainless steel mounting plate. The plate allows the meter to be hung on a
hook or on a nail. It also allows the meter to be permanently mounted with a screw.
The bottom part of the stainless steel mounting plate is covered with pressure
Note
sensitive adhesive. To install it onto the meter, peel off the pink colored releasing film;
1
press it firmly onto the back of the meter (see figure 6).
2
3
4
5
Figure 6. Left, using Velcro for mounti n g. Right, using stainless steel plate for mounting.
Copyright 2007-2016, Auber Instruments All Rights Reserved.
No part of this manual shall be copied, reproduced, or transmitted in any way without the prior,
written consent of Auber Ins truments. Auber Instruments retains the exclusive rights to all information
WWW.AUBERINS.COM
SET
SET
SET
SET
SET
SET
included in this document.
SET
SET
SET
SET
SET
P2/2