Normal operation

During normal operation the actuator is driven in an out using the appropriate buttons on the remote control. Situations
may arise which require a particular action. These situations are described below.
Low battery level
If the battery level goes below a warning level while operating the actuators, three short buzzer tones are sounded.
This means that the battery requires recharging, and the actuators can therefore only be driven in one direction
(normally in).
Flat battery
If the battery level goes below a minimum level while operating the actuators, one short buzzer tone is sounded. The
actuators will stop immediately and the controller will shut down. The battery must be charged. You must wait at least
one minute before inserting a new battery into the controller. If you are using actuators with position feedback (hall
effect), you will need to perform 0-point calibration. See below.
Extreme battery load
The battery may become too hot if it is subject to extreme load due to repeated operations in rapid succession, thereby
activating the battery's safety thermostat. The controller will shut down and the battery must be allowed to cool for
at least half an hour before being used again. You must wait at least one minute before inserting a different battery
into the controller. If you are using actuators with position feedback (hall effect), you will need to perform 0-point
calibration. See below.
0-point calibration
If you use actuators with position feedback (hall effect), you will need to perform 0-point calibration on the controller
so that the innermost position of the actuators can be recorded. This must be done for all actuators which use position
feedback. The controller will sound two short buzzer tones following the beep when the battery is inserted if 0-point
calibration is required.
It is also possible to initiate manual 0-point calibration by holding in the relevant 'in' buttons for 10 seconds. The
controller sounds a beep followed by 2 long buzzer tones and 0-point calibration can be performed by pressing the 'in'
button and holding it so that the given actuator(s) is/are driven in. Hold down the button until the actuators have been
driven fully in. The controller has now been 0-point calibrated and the system may be used normally. The actuators can
only be driven in during 0-point calibration.
|