3Shape TRIOS S1A-12 Manuel de configuration - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel de configuration pour {nom_de_la_catégorie} 3Shape TRIOS S1A-12. 3Shape TRIOS S1A-12 13 pages.

Step 5: After cleaning, disinfection, and sterilization the color calibration
kit must be reassembled before use. Ensure that all parts are dry before
reassembly. If not used immediately place the color calibration target in the
sleeve and leave the color calibration adaptor in the sterilization pouch.
IMPORTANT NOTICE
The calibration target is a delicate optical component.
During disassembly and assembly of the calibration kit
only the end-part to the calibration target (opposite of the
calibration patches) must be touched, and the target must
only be cleaned with air.
6.3 Cleaning and high-level disinfecting
other parts of the system
After patient treatment, clean and disinfect all other surfaces of the system
(except the scanner tip and color calibration kit, for which the previous
sections apply). Also note that the optical window is a delicate optical
component and should be handled with extra care during cleaning and
disinfection (see below).
For general cleaning, a soft cloth and a mild detergent can be used.
Recommended disinfection solution:
Denatured alcohol (IPA – Isopropyl Alcohol or Ethanol) – typically 60-70% Vol.
To disinfect:
Step 1: Apply the disinfection solution to a soft, lint-free, non-abrasive
cloth; the cloth must only be damp.
Step 2: Wipe the surfaces with the damp cloth.
Step 3 (optical window on scanner only): Dry the optical window with
another clean, soft, lint-free, non-abrasive cloth.
CAUTION
The scanner's optical window is a delicate optical
component. Its clean and undamaged surface is critical for
the scan quality. The window should be handled with care.
Make sure it is free from lint, stains, and other kinds of dirt.
CAUTION
Never submerge the TRIOS scanner
into the cleaning solution.
CAUTION
Do not use ammonia-based or chloride-based solutions,
or acetone on any surface. Before returning the scanner
to its holder, ensure both are dry.
CAUTION
Do not spray directly onto any parts of the system,
as the liquid solution can accumulate at the seals
and leak into the scanner, monitor, PC, cords, or pod.
IMPORTANT NOTICE
The scanner shell might get yellow-toned over
a long period of use.
IMPORTANT NOTICE
All surfaces that the protection tip comes in contact with,
particularly the scanner tube and its optical window, must
always be cleaned and disinfected before attaching the
protection tip.
6.4 Additional cleaning in case of poor scan quality
During scanning, the tip mirror may become unclean, resulting in a loss of
scan quality. In this situation, you can clean the mirror following these steps:
Step 1: Remove the scanner tip.
Step 2: Apply alcohol to a clean cloth or cotton-tipped swab and wipe
the tip mirror until it is clean. Use alcohol that is free of impurities, as
those may stain the mirror. Ethanol or propanol (ethyl/propyl alcohol)
can be used.
Step 3: Wipe the tip mirror dry with a dry, lint-free cloth to avoid leaving
dust and fiber on the mirror.
CAUTION
Do not use acetone or any oxidizing solutions to clean
the tip mirror.
If you still see problems, check the scanner's optical window for cleanliness,
and apply the same procedure if necessary.

7. Disposal

Described devices follow the Waste Electronical and Electronic
Equipment Directive. The Following symbol on the product,
the accessories or packaging indicates that this device
must not be treated as unsorted municipal waste, but must
be collected separately!  Please contact your 3Shape reseller,
who can inform you about the local requirements for disposal
of electronical, clinical waste. By disposing of the device
in the proper manner, you help to avoid possible hazards
for the environment and public health that could otherwise
be caused by improper treatment of waste equipment.
7.1 Disposal of the scanner tip
Sterilize the tip as described above. The scanner tip can then be disposed
normally as other clinical waste.
CAUTION
Dispose the scanner tip as other clinical waste.
CAUTION
The scanner tip must be sterilized before disposal.
7.2 Disposal of the TRIOS IOS system
For disposal of the TRIOS IOS system, please contact your 3Shape reseller, who
can inform you about the local requirements for disposal of electronic waste.
11