3Shape TRIOS S1P-4 Manuel de configuration - Page 12
Parcourez en ligne ou téléchargez le pdf Manuel de configuration pour {nom_de_la_catégorie} 3Shape TRIOS S1P-4. 3Shape TRIOS S1P-4 13 pages.
8. Technical Maintenance
CAUTION
All technical service and repair of the TRIOS IOS system
must be performed by 3Shape or 3Shape-authorized
personnel only.
The user doesn't need to perform any regular maintenance other than
calibration, cleaning, disinfection and sterilization, as described here.
Preventive inspections and scheduled maintenance are not required.
System S1A (MOVE/MOVE+ cart) only
Changing the fuses (Figure 2, S20):
Shut down the MOVE/MOVE+ cart by following the procedure described
in Setting up & Starting TRIOS for the first time and disconnect the mains
power cable from the MOVE/MOVE+ cart.
CAUTION
The mains power cable must be disconnected before
changing fuses.
Using a pair of tweezers or similar, push in the two tabs on each side of the
fuse drawer simultaneously, as shown in the subfigure S20.
The drawer will begin to emerge from its slot.
It is then possible to remove the drawer and replace the two fuses. Two spare
fuses are provided with the system on delivery.
WARNING
Only fuses with the correct rating and type
(250 V, T 3.15 A H) must be used.
To replace the fuses, first mount them in the drawer, then insert the drawer
back into the slot and push all the way in until the tabs click back into place
and the drawer is no longer loose in the slot. Do not attempt to insert
the fuses directly into the slot on the MOVE/MOVE+ cart.
The mains cable can now be reconnected, and the TRIOS IOS system started up.
Changing the dongle for the MOVE/MOVE+ cart (software license
protection):
The USB wireless dongle, the USB license dongle and the USB wifi dongle,
and the Bluetooth dongle are located behind the plastic lid on top of the
backpack of the MOVE/MOVE+ cart. You can remove the lid by opening
the screws holding the lid, as indicated by the arrows in figure S21.
WARNING
The scanner must not be operated while the ventilation
grid on the back side of the MOVE/MOVE+ cart is removed.
9. Consumable Accessories
The consumable accessories to the TRIOS IOS system are the scanner tips
and the calibration kit.
When the scanner tips are disposed, new tips can be ordered by contacting
your 3Shape reseller. Also, a new calibration
kit can be ordered from your 3Shape reseller.
12
10. Vigilance
In case of any serious incident that has occurred in relation to the use of the
device, please report the incident to 3Shape at email: [email protected]
and you can also report it to the competent authority of the state in which
the user and/or patient is established.