Plié Real Time Data Display
The Plié Real Time Data Display allows the viewing of live data as the user ambulates. The Plié Real Time Data
Display provides a data rate of 4 Hz (4 samples per second). Due to the limited data stream bandwidth (knee
microprocessor data rate is 1000 Hz), the Plié Real Time Data Display should not be used for setting the userʼs
computer parameters. To access the Plié Real Time Data Display, select the Live Data icon in the Plié Control
main window. To stream live data, select the Stream Data icon. The Plié Real Time Data Display is illustrated in
Figure C.
!
The Plié Real Time Data Display is useful for verifying calibration of the sensors. If the user unweights
Tip
!
the knee, the Moment trace should have zero value. If the knee is fully extended, the Angle should have
!
zero value.
Stability Parameters
The Stability Parameters window provides access to advanced stumble recovery settings via two tabs. To
optimize the advanced stumble recovery settings, follow the steps provided in the Plié Setup Wizard. To access
the Stability Parameters window, select the More Params icon in the Plié Control main window. The Stability
Parameters window is illustrated in Figure D.
Manually adjusting the Stability Parameters is not recommended. If the Plie I Mode box is checked, the
advanced stumble recovery is not active. The knee should only be in Plie I Mode during set-up.
Plié® 2.0 MPC Knee Instructions for Use (R-720-109 Rev. D), English!