Audiomatica CLIO FW-02 Manuel d'installation - Page 14

Parcourez en ligne ou téléchargez le pdf Manuel d'installation pour {nom_de_la_catégorie} Audiomatica CLIO FW-02. Audiomatica CLIO FW-02 15 pages.

Audiomatica CLIO FW-02 Manuel d'installation
To ensure a 100% correct calibration you also need to inspect the phase responses
of both measurements. To do this press the phase button
and verify that you
obtain a straight line (red curves) the readings in this case should be around zero
degrees in both cases.
As a final test click on the In-Out Loop
button for channel A; then click on the
In-Out Loop
button for channel B; then be sure that both input sensitivity are at
0dBV; click on the generator icon
to play the 1kHz sinusoid; at last press the F4
keystroke to invoke again the Multi-Meter; the expected result is shown now.
14
3 CLIO INSTALLATION