Calibration Traceability:
Apogee SI series infrared radiometers are calibrated to the temperature of a custom blackbody cone
held at multiple fixed temperatures over a range of radiometer (detector/sensor body) temperatures. The
temperature of the blackbody cone is measured with replicate precision thermistors thermally bonded to
the cone surface. The precision thermistors are calibrated for absolute temperature measurement against
a platinum resistance thermometer (PRT) in a constant temperature bath. The PRT calibration is directly
traceable to the NIST.
Spectral Response:
Spectral response of SI series
infrared radiometers. Spectral
response (green line) is
determined by the germanium
filter and corresponds closely to
the atmospheric window of 8-14
µm, minimizing interference
from atmospheric
absorption/emission bands
(blue line) below 8 µm and
above 14 µm. Typical terrestrial
surfaces have temperatures that
yield maximum radiation
emission within the atmospheric
window, as shown by the
blackbody curve for a radiator at
28 C (red line).
7