TECHNICAL DATA
HANDSWITCH DISPLAY
ERROR CODE
HOT
E00
E01
E02
E12
E13
E14
E24
E25
E26
E36
E37
E38
E48
E49
E55
E56
E60
E61
If it is not possible to resolve an error as described, disconnect the power cord, wait a few minutes and reset the table again.
If the error still occurs, remove the table from the power source and contact customer service.
(Error Code list is valid from firmware 1.7.5 and higher)
18 |
SOURCE
The control box monitors the duty
cycle (time-controlled) and its max
temperature. A value has been exceeded.
M1 – Internal Fault
M2 – Internal Fault
M3 – Internal Fault
M1 – Defect
M2 – Defect
M3 – Defect
M1 – Overcurrent
M2 – Overcurrent
M3 – Overcurrent
M1 – Not connected
M2 – Not connected
M3 – Not connected
Overcurrent drive 1
Overcurrent drive 2
Drive 1 - Synchronization is lost
Drive 2 - Synchronization is lost
Collision identified
Drive replaced
Wait until the "HOT" display goes out - the table is working properly again
Unplug and contact customer service
Unplug.
Correct the external short circuit and/or verify the cable to the drives.
Plug cables into the control box and re-run table.
Collision
resolve if necessary
Max load exceeded
remove excess load from the table
If the problem still exists, please contact customer service
Motor not connected.
Check cable / connector to the drive.
Reset the table (see page 15).
Collision
resolve if necessary
Max load exceeded
remove excess load from the table
If necessary, reduce the load on the table
Reset the table (see page 15).
Option ISP function. Remove source.
The connection to the drive is interrupted or a new drive is connected.
Reset the table (see page 15).
SOLUTION