Dynamic TEACH and Manual Adjust
Moves switching threshold value up or down to make adjustments
Slide Mode switch to ADJ to enter Adjust mode
Press "+" to increase; press "-" to decrease
- GREEN display shows the switching threshold value
- 2 seconds after adjustment, GREEN display will flash 3 times to confirm
Slide Mode switch to RUN to complete operation
Remember: Manual adjustments are disabled when Auto Thresholds are ON
Follow these steps to perform a Dynamic TEACH:
NOTE: TEACH Selection must be programmed to dYn tcH (see
1. Enter Adjust Mode.
Method
SET Button 5
Remote Input 6
2. Enter Dynamic TEACH.
Method
SET Button
Remote Input
3. Present ON and OFF Conditions.
Method
SET Button
Remote Input
SET Button: 0.04 seconds ” "Click" ” 0.8 seconds
5
Remote Input: 0.04 seconds ” T ” 0.8 seconds
6
Action
Set Mode switch to ADJ
No action required; sensor is ready for
Dynamic TEACH method
Action
Click the SET Navigation key
Single-pulse remote input
Action
Present ON and OFF conditions
Present ON and OFF conditions
Program
Mode)
Display: Red - Signal Level;
Display: Flashes "dYn tch" then holds
on "1234 dYn"
T
Display: Red - Signal Level;
7
Result
Green - Threshold
Result
Result
Green - Threshold