3.8

Printer setting

Please start as described in the beginning of chapter 3 to enter this menu and „Prt_" is shown
in the display.
Setting has to be "Prt_0", because scale has no interface for connecting PC or printer.
SP – x.x
Version number
„xxxxxx"
Software number
"CAPxx.0"
Capacity / Maximum load of the scale
Scale is overloaded. The current weighing is higher than the valid
„nnnnnn"
capacity.
"------"
Data is unstable
"LLLLLL"
Output of load cell is lower.
"HHHHHH"
Output of load cell is higher or turn on zero is higher than range.
- Lo -
Low battery – Recharge the rechargeable battery immediately.
FULL
Rechargeable battery has been charged fully.
"_Inlt_"
Faulty weighing
Fault
Faulty weighing result
Display shows „HHHHHHHH"
Display shows „LLLLLLLL"
Rear display doesn't show
anything
Rechargeable battery is not
loading although loading is
shown
Rechargeable battery is not
loading and no loading is
shown
Quickly discharging of
rechargeable battery
Display shows „_InIt_"
9
4.

Display Symbols

5.
Display Symbols
Cause
Parameter of load cells have
changed
Over zero range
Load cell or PCB damaged
Load cell or PCB damaged
Parameter fault
Load cell damaged
Rechargeable battery is
damaged
Charging unit for
rechargeable battery is
damaged
Old rechargeable battery
Faulty weighing
Solution
Re-calibration
Re-calibration
Correction of setting as
shown in the manual
Replace rechargeable battery
Replace rechargeable battery
Re-start the scale or Re-
calibration
KWE-190107-Rev008-UM-en