DATATEC Y.I.C. TECHNOLOGIES EMProbe Manuel de configuration - Page 4
Parcourez en ligne ou téléchargez le pdf Manuel de configuration pour {nom_de_la_catégorie} DATATEC Y.I.C. TECHNOLOGIES EMProbe. DATATEC Y.I.C. TECHNOLOGIES EMProbe 7 pages.
Installation and Hardware Setup
To safely and accurately setup the EMProbe solution, please refer closely to the following
information regarding the full installation process.
1. Install communication software for the spectrum analyzer: You must determine
which software to install from the manufacturer's website. For SCPI controlled spectrum
analyzers, any software which installs NI-VISA can be used. This includes NI MAX,
Keysight Connection Expert and RIGOL UltraSigma. It is recommended to install one of
the following:
a.
NI-MAX
Recommended for most SCPI-controlled spectrum analyzers. Download NI System
Configuration from the National Instruments website: www.ni.com. When "Additional items
you may wish to install" is shown, you must have at least NI Measurement & Automation
Explorer selected. Other items do not affect the EMViewer runtime and can be enabled at the
user's discretion.
b.
Keysight Connection Expert
Recommended to use with Keysight spectrum analyzers.
Go to
www.keysight.com
website to download the latest version. On the website, click on the 'Download' button and
follow the steps as instructed.
4
Sentinel House, Ancells Business Park, Harvest Crescent, Fleet, Hampshire, GU51 2UZ, UK
and search for 'IO Libraries Suite', or
Copyright © Y.I.C. Technologies Ltd
All rights reserved.
click here
to open the