Cooey B61T Manuel de l'utilisateur - Page 13
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Cooey B61T. Cooey B61T 19 pages.
Humidity: 10% to 93% RH
Pressure altitude: 70KPa~ 106Kpa
CLASSIFICATION
1. Internally powered equipment; (supplied by Built-in Lithium Ion Polymer Battery )
2. Type BF applied part;
3. IP21;
4. No Sterilize requirement;
5. Not category AP / APG equipment
6. Mode of operation: continuous
EMC Declaration
Guidance and manufacturer's declaration – electromagnetic immunity
The "Digital Blood Pressure Monitor B61T" is intended for use in the electromagnetic environment specified below. The customer
or the user of the "Digital Blood Pressure Monitor B61T" should ensure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
IEC 60601
test level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV differential
mode
2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
Version Number: 1.0
Compliance
level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV differential
mode
2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
"Digital Blood Pressure Monitor B61T" requires
continued operation during power
mains interruptions, it is recommended
that the "Digital Blood Pressure Monitor B61T" be
powered from an uninterruptible power
supply or a battery.
–