3-mm Lightguide - For use when measuring energy at the end of a 3-mm lightguide.
5-mm Lightguide - For use when measuring energy at the end of a 5-mm lightguide.
8-mm Lightguide - For use when measuring energy at the end of an 8-mm lightguide.
Figure 9. Lightguide Source Screens (Shown in Peak Intensity Mode)
10. Press and release the Mode button to select the operating mode. Repeatedly press the Mode button
until the mode you desire appears on the second line of the display. The different operating modes are:
Peak Intensity - Shows the highest intensity in mW/cm
measurement and stores that value in the display.
Intensity - Shows the intensity in mW/cm
readout of intensity.
Dose - Shows the total energy dose in mJ/cm
Note: This measurement is time dependent.
11. Turn on the spot lamp to begin testing.
Dymax ACCU-CAL™ 50-LED Radiometer User Guide
2
seen by the detector during the
2
at the detector during the measurement. This is a dynamic
2
at the detector during the measurement.
11