Product Cleaning and Care

Product cleaning is limited to wiping the product with a damp cloth. Do not soak.
Isopropanol Alcohol or household cleaners may be used for cleaning the product.
Always inspect the quartz window for cleanliness before use. Foreign material can
cause permanent damage to the window. Clean with Isopropanol Alcohol to
remove smudges or foreign material. Damaged or permanently etched windows
should be replaced.
Do not use compressed air to removed particle debris inside the emitter as it may
damage the high-speed cooling fan.

Glass Window Replacement

1. Before starting, disconnect power to the unit.
2. Lay the emitter on its side with the cover facing upwards. Using a 5/64" hex key,
loosen and fully remove the two screws (Figure 4).
3. Remove the side cover by lifting from the bottom as shown in Figure 5.
Figure 4.
Remove Screws
4. Slide the damaged glass cover out (Figure 6), taking care not to touch the metallic
reflector surfaces.
NOTE: If the old window has shattered, be sure to remove all debris before
attempting to install the new glass window.
5. Unwrap the new window and carefully clean with the enclosed alcohol swab.
Holding the window by its sides, slide into the groove (Figure 7). Take care not to
touch the glass with bare hands, as any residue left on the window can adversely
affect the performance of the unit.
1 0
Figure 5.
Remove Side Cover
U S E R G U I D E
|