Active Motif 53081 Manuel - Page 12

Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} Active Motif 53081. Active Motif 53081 15 pages. Cooled sonication platform

Appendix

Section A. Attaching the Alternate Spring Bracket

The alternate spring bracket was provided in case you need to remove the converter/probe
bracket from the top of the rod, which you would need to do if your converter/probe is too large
to be held by the bracket. The alternate bracket enables you to reattach the continuous force
spring so that it can support the weight of the platform when it is moved up and down.
To replace the converter/probe bracket with the
alternate spring bracket, the rod and platform will
need to be removed from the sound enclosure. To
do this, lay the sound enclosure box flat with the
door facing up and open.
Loosen the screw on the top of the box that holds
the rod in place using the supplied hex wrenches
(top image). Then remove the three screws and
bracket supporting the rod on the bottom of the
sound enclosure (bottom image). Gently slide the
rod down 1-2 inches through the hole in the bottom
of the sound enclosure to bring the top of the rod
below the interior insulation. Then angle the top
of the rod into the probe chimney and gently slide
the rod up until the entire platform is released from
the sound enclosure. With the platform removed
from the sound enclosure, follow the instructions to
replace the converter/probe bracket.
www.activemotif.com
9