4.
REPROCESSING
RACCOMANDATION: reprocess shall be performed within 30 minutes of the end of use.
The micromotors have been tested for 250 cycles (as required by EN ISO 14457).
Before sanitisation operations, the operator must detach the handpiece from the micromotor and the micromotor from the dental
units tubing.
Before reprocessing, visually inspect the micromotor. Reprocess only micromotors free of any unacceptable deterioration such as
corrosion and damages.
To maintain a level of hygienic security, at the end of each use and within a short time he must clean, disinfect and sterilize the
micromotor.
4.1. CLEANING
For cleaning follow the instructions below:
• Create a cleaning solution following the instructions for light organic debris.
Use a cleaner that is legally marketed and approved in your country (see paragraph PRODUCTS TO BE USED).
• Soak a lint-free cloth in the prepared detergent solution and tightly wrap the micromotor for 2 minutes.
Pay particular attention to have the cloth contact the front and rear connections of the device.
• Unwrap the device and dispose of the wipe.
• Soak a new lint-free cloth in the detergent solution and wipe the device by the following:
1
Wipe the external surface. Repeat this step using a fresh portion of the wipe if any visible soil remains.
2
Wipe the mating surface where the handpiece connects by wrapping the shaft housing with the wipe and rotating the device. Repeat this step with
a fresh portion of the wipe if visible soil remains.
3
Wipe the edge of the anterior mating surface. Press the wipe against the surface and rotate the device a few times both clockwise and
counterclockwise. Repeat this step with a fresh portion of the wipe if visible soil is observed.
Pay particular attention to the screw heads, recesses, and crevices of the device.
4
Wipe the edge of the rear mating surface of the device. Press the wipe against the surface of the device and rotate the device.
5
Repeat only step 4.1.
• Soak a new lint-free cloth in water (do not use saline solution) and wipe the micromotor by the following:
1
Wipe the external surface.
2
Wipe the mating surface between the handpiece and micromotor by wiping the stud and rotating the device a few times clockwise and
counterclockwise.
3
Wipe the edge of the anterior mating surface. Press the wipe against the surface to achieve good contact and rotate the device a few times both
clockwise and anti-clockwise.
4
Wipe the edge of the rear mating surface of the device. Press the wipe against the surface to have good contact and rotate the device a few
times.
• Using compressed air, dry the article.
• Using a new clean lint-free cloth, wipe the device to remove any excess of water or lubricant.
VISUAL INSPECTION.
At the end of the cleaning procedure, visually inspect the device under normal lighting conditions.
If any visible soil is present, repeat the cleaning procedure.
I-MMr • i-MMr L • i-MMs
EN
7